The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended for potentially
infectious specimens in Protection of Laboratory Workers from Infectious
Disease Transmitted by Blood, Body Fluids, and Tissue, 2nd edition; Approved
Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user
protection and it can be used as reference material for instructions on laboratory
safety.
When performing this procedure, do not use force on the syringe at any time.
1. At the Utilities menu, select
2. Disconnect the waste tubing
V24. This is the white fitting at V24
3. Fill the cleaning syringe
EZ KLEEN and attach it to
the white fitting. Refill the
syringe EZ KLEEN as needed
during the following steps.
4. Disconnect the sample tubing
(5)
syringe
sample tubing into a beaker to
catch fluid as you backflush
the flowcell
5. Gently push the plunger on
the cleaning syringe to flush
the EZ KLEEN through the
Perox flowcell
with deionized water.
6. Reattach the sample tubing
to the top of the sample
syringe
7. Refill the syringe with EZ KLEEN.
8. Disconnect the shuttle line
sample tubing into a beaker to catch fluid as you backflush the flowcell.
9. Gently push the plunger on the cleaning syringe to flush the EZ KLEEN
through the Perox flowcell
10. Reattach the shuttle line
11. Refill the syringe with EZ KLEEN.
5-22
CAUTION
(4)
from the top of the sample
(6)
and place the
. Repeat
(2)
(6)
(7)
to open the Exerciser tab.
Exerciser
that comes from the Perox flowcell
(1)
(3)
.
4
V25
(5)
(7)
from the side of the UFC
. Repeat with deionized water.
(2)
to the side of the UFC
at
(2)
1
2
7
3
5
V6
V24
V15
8
(8)
and place the
(8)
.
Troubleshooting the Analyzer
V8
V5
V7
V18
V16
V17