Autosampler - Flash Memory Failed, Reset Autosampler
The autosampler flash memory failed to respond to a routine command and might
be defective. An autosampler reset is required.
Corrective Action
1
Select
2
Wait about 1 minute and select On to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Initialization Required
In diagnostic mode, the analyzer sent a position command before the appropriate
initialization was performed. For example, a position mixer command could have
been sent before a mixer initialization.
Corrective Action
Call Siemens Service for assistance.
Autosampler - Input Queue Jammed
Possible Cause
1
Rack has
jammed the
leading edge of
input-queue
blade.
2
Queue pan
motion is
impeded.
3
Feed motor
failed.
Autosampler - Input Queue Motion Denied
To prevent the collision of autosampler components, the system did not execute
the command to move the autosampler input queue.
Autosampler - Input Queue Motion Denied, Reset Autosampler
The autosampler attempted an inappropriate input queue motion. An autosampler
reset is required.
Status Line messages
Off
on the analyzer touchpad.
Corrective Action
Remove and reset rack.
Check for foreign objects in input-queue pan.
Call Siemens Service for assistance.
7-23