Anisocytosis (Aniso) - Siemens ADVIA 2120 Operator's Manual

Table of Contents
Flag
Triggering Criteria
LPLT
%LPLT > 10% PLT
LS
BASO d/D < 0.15 and
%NEUT
MACRO
%MACRO
MICRO
%MICRO
MPO-D
[%PMN − (%NEUT +
(MO)
%EOS)] ≥ 25, no NRBC
flag and a valid MN-PMN
valley (d/D ≥ 0.15)
NRBC
1. WBCu ≥ to 3000 cells
/ µL and % nRBC ≥ 2.0%
(# nRBC / 100WBC)
or
2. # nRBC≥ 200 cells / µL
PLT-CLM
Clumps Count > 150
(NW)
RBCF
RBCF > 100,000 cells/µL
RBCG
RBCG > 100,000 cells/µL

Anisocytosis (ANISO)

Definition
The Anisocytosis flag is triggered if variation in RBC volume (RDW) is equal to
or greater than 16%. The Red Cell Distribution Width (RDW) parameter is the
coefficient of variation of the cell volume distribution on the RBC volume
histogram.
Default trigger values for the three severity levels are:
+
++
+++
6-4
CBC
30%
2.5%
2.5%
RDW = 16.0% to 17.9%
RDW = 18.0% to 22.0%
RDW > 22.0%
CBC /
CBC / DIFF
DIFF
/ RETIC
CBC /
RETIC
RETIC
Flags