Backflushing The Perox Flowcell And Shuttle And Reaction Chambers - Siemens ADVIA 2120 Operator's Manual

Table of Contents
Shuttle line from UFC assembly (
Sample line from syringe (
Sheath line from hydraulic valve (
V15 - perox or V19 - RBC
Cleaning the Flowcells off the Analyzer: Step 7
Checking Analyzer Performance
IMPORTANT
An inspection scope must be available for this procedure. Do not remove the
flowcells unless you are trained in how to align and focus them.
1. Bring the analyzer to the
2. At the Operations menu, select
background count.
3. Check the flowcell connections for leaks.
4. Check the focus and alignment of the flowcell.
5. Check the gains.
6. Run a whole-blood primer, then run controls to verify analyzer performance.
7. If the control results are acceptable to the laboratory, no additional action is
required and normal operation can be resumed.
Backflushing the Perox Flowcell and Shuttle and Reaction
Chambers
Materials required:
Flowcell cleaning kit, PN 113-B711-01
EZ KLEENBeaker
Deionized water
Analyzer mode: Ready to Run
All products or objects that come in contact with human or animal body fluids
should be handled, before and after cleaning, as if capable of transmitting
infectious diseases. Wear facial protection, gloves, and protective clothing.
Troubleshooting the Analyzer
BIOHAZARD
)
3
)
2
)
1
mode.
Ready to Run
, select
Startup
Refresh
, and verify the
5-21