Event Text
Autoslide Command Error
Autoslide Daily Shutdown
Incomplete
Autoslide Daily Startup
Incomplete
Autoslide is Processing
Slides
Autoslide Manual Inlet
Door Cannot be Opened
Autoslide Manual Inlet
Door Failed to Open
Autoslide operation
cannot proceed as
Compressor is Not Ready
Autoslide Rejects
Command Execution
Autoslide Weekly
Shutdown Incomplete
Autoslide Weekly Startup
Incomplete
Autoslide Weekly Startup
Needed
ADVIA Autoslide Slide Maker Stainer
Severity
Probable Cause
Warning
Autoslide command error, or
EEPROM failed.
Warning
An error occurred that prevented
the autoslide daily shutdown from
completing.
Warning
An error occurred that prevented
the autoslide daily startup from
completing.
Warning
Attempt was made to go to a
configuration screen while slides
were being processed.
Warning
The manual inlet door cannot be
opened because the stainer is not
enabled for use or the ADVIA
2120/2120i Autosampler is active.
Warning
An error occurred while attempting
to open the manual inlet door.
Warning
Instrument is in standby, so
compressor is off.
Warning
The Autoslide is unable to execute
the command at this time.
Warning
An error occurred during the
weekly shutdown.
Warning
An error occurred during the
weekly startup.
Warning
Weekly startup is now required.
Suggested Action
Retry the command.
Check message log for errors, correct
any problems, and retry the daily
shutdown. If still not successful,
perform a Mechanical Initialization and
then retry the daily shutdown.
Check message log for errors, correct
any problems, and retry the daily
startup. If still not successful, perform a
Mechanical Initialization and then retry
the daily startup
Wait until all slides are completed
before attempting to change
configuration.
Enable the stainer in the Autoslide
options menu or stop Autosampler if
appropriate, and try again.
Check for something blocking the Stat
door, and try again.
Take instrument out of Standby mode.
Action depends on the situation.
Generally no action is required.
Check message log for other errors,
correct problem, reset Autoslide, and
repeat weekly shutdown.
Check message log for other errors,
correct problem, reset Autoslide, and
repeat weekly startup.
Perform weekly startup.
10-51