Replacing The Perox Lamp - Siemens ADVIA 2120 Operator's Manual

Table of Contents
b. Visually check for bubbles in the open-tube sample line.
c. Run enough controls to verify analyzer performance. If control results

Replacing the Perox Lamp

To prevent injury to eyes from shattered glass, wear safety glasses when handling
tungsten-halogen lamps.
To avoid severe burns, allow sufficient time for the lamp to cool. After the
analyzer is turned off, wait at least 10 minutes.
Fingerprints on the lamp could cause severe degradation in lamp performance.
DO NOT touch the lamp with your fingers.
Material required: Replacement lamp, PN 065-B075-01
Time: Replacement - 20 minutes
Analyzer mode: Off, wait at least 10 minutes.
To avoid exposure to shock hazards and/or damage to the instrument while
performing this procedure, power off the analyzer before proceeding.
To replace the perox lamp (
IMPORTANT
Make sure that the power is off before disconnecting the lamp.
1. Disconnect the lamp power connector (
assembly by squeezing the sides of the red connector and pulling it apart.
5-38
are not acceptable, check that the probe has been properly installed. If no
problem is found, recalibrate the affected channel.
WARNING
CAUTION
Adjustment and checkout - 20 minutes
ELECTRICAL WARNING
)
1
) in the back of the perox optics
2
FRONT OF ANALYZER
Troubleshooting the Analyzer