Corrective Action
Using the Customer Parameters window, define the parameter, Max Number of
Automatically Proposed Reruns, to specify the maximum number of reruns
proposed by the Data Manager.
NULL - Status Modified
The status of the specified sample has been modified.
Corrective Action
No operator action required.
O
Open Port Error
A software preparation (initialization) error has occurred.
Corrective Action
1
Select
2
Select
software.
3
If the error recurs, select
Port
4
Verify that the Inst1 device is listed as ADVIA 2120/2120i and correct it if
necessary.
5
Check the communications port where the problem occurs (COM 1 or
MCA Serial board).
If the system is using the Host Spec 79 Network connection, Siemens Field
Service personnel should make sure the Host name and IP address are specified
correctly:
1
Use Windows Explorer to open the directory
C:\WINNT\system32\drivers\etc
2
Double-select
3
In the Open with box, select
4
At the bottom of the document, verify the host IP address and the host
computer name match those supplied by the host computer administrator.
Make sure the address and name are separated by at least one space.
5
If there is an error in the Host name or IP address, correct it. Then on the
File menu, select
6
Close Notepad.
7-66
at the Log On/Off window.
Shut Down ADVIA
CTRL+ALT+DELETE
Customize
Configuration.
Hosts
.
.
Save
and log back onto the system to restart the
menu,
System Setup
.
Notepad
and select
OK
,
,
Tools Modify
.
Status Line messages