Running
Displayed after pressing the autosampler Start/Stop button or during sample
processing.
Possible Cause
If the system halts
in the Running
state, there could
be a hardware or
software failure.
S
# $ > Sample ID Not Read - Alarm
Sample-tube barcode label is unreadable or missing. The number of consecutive
occurrences of this condition has met the criterion specified for an alarm
message.
Possible Cause
1
Barcode label
improperly
placed,
obscured, or
not present.
2
Barcode reader
faulty.
Sample ID Not Read - Stop
Sample-tube barcode label is unreadable or missing. The number of consecutive
occurrences of this condition has met the criterion specified for a stop message,
and the autosampler has halted.
Possible Cause
1
Barcode label
improperly
Status Line messages
Corrective Action
a
Reset the analyzer by selecting
Indicators, Analyzer Reset
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-aspirate
light is still off, follow these steps to exit the ADVIA
2120/2120i software, turn off the analyzer and then
restart.
i
Select
Shut Down ADVIA
window.
ii
Select
at the analyzer touchpad.
Off
iii
Select
CTRL+ALT+DELETE
the system to restart the software.
iv
Restart the analyzer by selecting
analyzer touchpad.
Corrective Action
Verify that a barcode is present and correctly applied to
the sample tube (with the label flush against the rim of
the stopper). Verify that the barcode is clean and of
specified quality.
Call Siemens Service for assistance.
Corrective Action
Verify that a barcode is present and correctly applied to
the sample tube (with the label flush against the rim of
Utilities, Exerciser,
.
at the Log On/Off
and log back onto
at the