If the analyzer is locked in the standby state, verify that the pressure and vacuum
readings are within range and adjust if necessary. Select Standby to exit.
Status BTRIEVE - Alarm Not Defined
An alarm sent from the analyzer is unknown to the Data Manager (not defined in
the Alarm Dictionary).
Corrective Action
See Alarm Dictionary.
Status BTRIEVE - Test Not Found
A test defined in the Control Dictionary has been erased from the Test
Dictionary.
Corrective Action
Enter test information in the Test Dictionary.
Syntax Error in AlarmExc.par
The parameter file, AlarmExc.par, is not correctly defined, or there is a blank line
in the file. No exclusive analytical alarm will be taken into account in the patient
QC calculation.
Corrective Action
Call Siemens Service for assistance.
System Preparation Failed
An error occurred during the initial system preparation phase.
Possible Cause
1
Software
communication
failure
2
Hardware error
3
Faulty CPU
board
7-92
Corrective Action
Follow these steps to exit the ADVIA 2120/2120i
software and then restart.
a
Select
Shut Down ADVIA
window.
b
Select
CTRL+ALT+DELETE
system to restart the software.
Refer to associated hardware error message for
appropriate corrective action.
Call Siemens Service for assistance.
in the Log On/Off
and log back onto the
Status Line messages