Valve Node 1 Error - Reset Required
Potential software communication error or a hardware failure associated with the
Valve Driver 1 Board.
Possible Cause
1
Hardware
failure
2
Faulty power
supply, fuse,
cable, Valve
Driver 1 Board,
or Can Bus
Scrambler
Board
Valve Node 2 Error - Reset Required
Potential software communication error or a hardware failure associated with the
Valve Driver 2 Board.
Possible Cause
1
Hardware
failure
Status Line messages
Corrective Action
a
Reset the analyzer by selecting
Indicators, Analyzer Reset
b
After about 1 minute, if the status line does not display
Ready to Run and the green ready-to-aspirate light is
still off, follow these steps to exit the ADVIA
2120/2120i software, turn off the analyzer and then
restart.
i
Select
Shut Down ADVIA
window.
ii
Select
at the analyzer touchpad.
Off
iii
Select
CTRL+ALT+DELETE
system to restart the software.
iv
Restart the analyzer by pressing
touchpad.
Call Siemens Service for assistance.
Corrective Action
a
Reset the analyzer by selecting
Indicators, Analyzer Reset
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-aspirate
light is still off, follow these steps to exit the ADVIA
2120/2120i software, turn off the analyzer and then
restart.
i
Select
Shut Down ADVIA
window.
ii
Select
at the analyzer touchpad.
Off
iii
Select
CTRL+ALT+DELETE
the system to restart the software.
Utilities, Exerciser,
.
at the Log On/Off
and log back onto the
at the analyzer
On
Utilities, Exerciser,
.
in the Log On/Off
and log back onto
7-97