Error.
2
40 PSI line is
out of range.
3
Needle is
bent.
4
Needle sensor
failed.
Autosampler - Needle Jammed
The needle failed to extend or retract during sample aspiration.
Possible Cause
1
40 PSI line out
of range.
2
Needle sensor
failed.
Autosampler - Needle Motion Denied, Reset Autosampler
The autosampler attempted an inappropriate needle motion. An autosampler reset
is required.
Corrective Action
7-30
will reset.
c
If the problem persists, select
touchpad.
d
Wait about 1 minute and select
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
a
Check pressure and adjust if necessary.
b
Check for pinched air lines supplying needle
cylinder.
c
At the analyzer touchpad, select Off.
d
Wait about 1 minute and select On to restart the
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
Replace needle.
Call Siemens Service for assistance.
Corrective Action
a
Check pressure and adjust if necessary.
b
Check for pinched air lines supplying needle
cylinder.
c
At the analyzer touchpad, select
d
Wait about 1 minute and select
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
Call Siemens Service for assistance.
Off
at the analyzer
to restart the
On
Off
.
On
to restart the