2
The cable connected to
the switch/indicator is
crimped.
3
Faulty power supply,
fuse, Switch/Indicator
board, cable, Parallel
Node Board, Can Bus
Scrambler Board, or
Switch/Indicator panel.
Initialization Failed - During Communication from Analyzer
Software communication error.
Corrective Action
Follow these steps to exit the ADVIA 2120/2120i software and then restart:
1
Select
2
Select
software.
If the problem persists, call Siemens Service for assistance.
Initialization Failed - During Configuration Download
Software communication error.
Corrective Action
Follow these steps to exit the ADVIA 2120/2120i software and then restart:
1
Select
2
Select
software.
If the problem persists, call Siemens Service for assistance.
Initialization Failed - During Download of Cycle Definition
Software communication error.
7-54
Select
crimped cable. Free the crimped cable if
necessary and restart the analyzer by selecting
On
Call Siemens Service for assistance.
Shut Down ADVIA
at the Log On/Off window.
CTRL+ALT+DELETE
at the Log On/Off window.
Shut Down ADVIA
CTRL+ALT+DELETE
i
Select
Shut Down ADVIA
On/Off window.
ii
Select
Off
at the analyzer touchpad.
iii
Press
CTRL+ALT+DELETE
onto the system to restart the software.
iv
Restart the analyzer by selecting
the analyzer touchpad.
Off
at the analyzer touchpad to check for
at the analyzer touchpad.
and log back onto the system to restart the
and log back onto the system to restart the
at the Log
and log back
at
On
Status Line messages