Listing The Pending Workorders; Running The Samples - Siemens ADVIA 2120 Operator's Manual

Table of Contents
Select
Select
4. Enter sample and patient information.
5. Use the
You must enter a sample ID number.
6. Select tests.
7. Select
The Access dialog box displays for the next workorder.

Listing the Pending Workorders

1. At the Customize menu, select
2. When the tools list appears, double select
3. Select the
4.
Optional:
created within a specific period of time.
5.
Optional:
sufficient.
6. Select

Running the Samples

Control Recommendations
Run controls in accordance with your laboratory protocol.
Run multilevel controls at the beginning of each shift.
Siemens recommends the use of ADVIA TESTpoint Hematology Controls
(Low, Normal, and High) and ADVIA TESTpoint Reticulocyte Control
(Low and High).
Run controls alone or at the start of the patient workload.
Optionally, you can run a retained patient sample periodically to monitor
performance trends.
Evaluate all control results before reporting patient results.
If control results fail to meet the laboratory's established criteria for
acceptability, you must evaluate all patient test results obtained in the
unacceptable run to determine if the patient results were adversely affected.
Perform and document appropriate corrective actions, which may include
recalibration and reassaying of patient samples, before reporting patient
results.
3-6
to make a new workorder for this patient.
Create
to create a new patient file for this name.
New Patient
key to move from field to field.
Tab
OK
to confirm the entries.
Pending
check box.
Use the To and From date boxes to restrict listing to workorders
Choose a report format from the Format list. DEFAULT is usually
to view the list, or select
Display
Tools View
.
File Management
to print it.
Print
.
Daily Routine
Table of Contents
loading