HGB Irregular Flow Rate - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
HGB Irregular Flow Rate - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The specified stop criterion is user definable in the
Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Stop Criteria is reset when the Start/Stop button is selected.
HGB Power Low - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
HGB Power Low - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
HGB Reagent Empty - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The reagent is empty, and the autosampler has
halted. The specified stop criterion is user definable in the Customize Menu,
System Setup, Alarm/Stop Criteria, Reagent Conditions.
Corrective Action
The stop counter is automatically reset after reagents are replenished and the
Reagent Installation window is updated. Go to Logs, Reagent Log, Reagent
Installation. Select Start/Stop to continue the run.