Autosampler - Input Shuttle Blocked
A rack in the input-shuttle position is preventing the autosampler from
completing a movement.
Corrective Action
1
Remove the rack from the shuttle position at the front of the input queue.
2
Select
Autosampler - Input Shuttle Not Empty
A rack in the input-shuttle position is preventing the autosampler from
completing a movement.
Corrective Action
1
Remove the rack from the shuttle position at the front of the input queue.
2
Select
Autosampler - Internal Communication Failure
During normal operation, the autosampler detected an internal, software-level
communication failure. This caused the autosampler to stop operation.
Corrective Action
1
At the analyzer touchpad, select
2
Wait about 1 minute and select
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Invalid Operation Mode, Reset Autosampler
The autosampler was unable to carry out a requested motion. The autosampler
might not be in the correct mode.
Corrective Action
1
At the analyzer touchpad, select
2
Wait about 1 minute and select
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Invalid Sequence Number
The autosampler sent the analyzer a sample ID sequence number that did not
match the expected sequence number.
Status Line messages
at the analyzer touchpad to restart autosampler operation.
Start
Start
at the analyzer touchpad to restart autosampler operation.
Off
.
to restart the analyzer.
On
Off
.
On
to restart the analyzer.
7-25