Basic Operation For A Thermocouple - Siemens S7-1200 System Manual

Hide thumbs Also See for S7-1200:
Table of Contents
Technical specifications
A.9 Thermocouple and RTD signal modules (SMs)
Table A- 168 Connector Pin Locations for SM 1231 AI 8 x TC bit (6ES7 231-5QF32-0XB0)
Note
Unused analog inputs should be shorted.
The thermocouple unused channels can be deactivated. No error will occur if an unused
channel is deactivated.
A.9.1.1

Basic operation for a thermocouple

Thermocouples are formed whenever two dissimilar metals are electrically bonded to each
other. A voltage is generated that is proportional to the junction temperature. This voltage is
small; one microvolt could represent many degrees. Measuring the voltage from a
thermocouple, compensating for extra junctions, and then linearizing the result forms the
basis of temperature measurement using thermocouples.
When you connect a thermocouple to the SM 1231 Thermocouple module, the two dissimilar
metal wires are attached to the module at the module signal connector. The place where the
two dissimilar wires are attached to each other forms the sensor thermocouple.
Two more thermocouples are formed where the two dissimilar wires are attached to the
signal connector. The connector temperature causes a voltage that adds to the voltage from
the sensor thermocouple. If this voltage is not corrected, then the temperature reported will
deviate from the sensor temperature.
Cold junction compensation is used to compensate for the connector thermocouple.
Thermocouple tables are based on a reference junction temperature, usually zero degrees
Celsius. The cold junction compensation compensates the connector to zero degrees
Celsius. The cold junction compensation restores the voltage added by the connector
thermocouples. The temperature of the module is measured internally, and then converted to
a value to be added to the sensor conversion. The corrected sensor conversion is then
linearized using the thermocouple tables.
For optimum operation of the cold junction compensation, the thermocouple module must be
located in a thermally stable environment. Slow variation (less than 0.1 °C/minute) in
ambient module temperature is correctly compensated within the module specifications. Air
movement across the module will also cause cold junction compensation errors.
If better cold junction error compensation is needed, an external iso-thermal terminal block
may be used. The thermocouple module provides for use of a 0 °C referenced or 50 °C
referenced terminal block.
922
Pin
X10 (gold)
1
L+ / 24 VDC
2
M / 24 VDC
3
GND
4
AI 0+ /TC
5
AI 0- /TC
6
AI 1+ /TC
7
AI 1- /TC
X11 (gold)
No connection
No connection
No connection
No connection
No connection
No connection
AI 2+ /TC
AI 4 I- /TC
AI 2- /TC
AI 4 I+ /TC
AI 3+ /TC
AI 5 M- /TC
AI 3- /TC
AI 5 M+ /TC
X12 (gold)
X13 (gold)
No connection
No connection
No connection
AI 6 I- /TC
AI 6 I+ /TC
AI 7 M- /TC
AI 7 M+ /TC
S7-1200 Programmable controller
System Manual, 03/2014, A5E02486680-AG
Table of Contents
loading

Table of Contents