GB091046, R
IRECTION
10-6-6
Isolated patient lead (source) leakage–lead to lead
Select and test each of the ECG lead positions (except ALL) on the LEAD selector, testing each to the
power and ground condition combinations found in:
•
Table 10-8 on page
•
Table 10-9 on page
Record the highest leakage current measured.
2
EVISION
10-16, or
10-17, as applicable.
Chapter 10 - Care & maintenance
VIVID E9 / VIVID E7 BT'13 S
M
ERVICE
ANUAL
10 - 23