GE Vivid E9 Service Manual page 252

With xdclear
Hide thumbs Also See for Vivid E9:
Table of Contents
GE
D
GB091046, R
IRECTION
5-10-2-1
General description (cont'd)
Power from the wall outlet (100 VAC to 230 VAC, 50/60 Hz) is connected to the Main Power Supply.
Figure 5-42 Main Power Supply block diagram
90 - 250 VAC 50/60 Hz MAINS
USB FROM BEP
TO MOTOR/BRAKE CONTROL
The Main Power Supply delivers the needed voltages to the rest of the system:
Internal Peripherals (115 VAC)
Front End Rack (DC power with several voltages)
-
-
-
-
Front End Rack (TXPSV1 and TXPSV2 for the transmitters)
Front End Rack (PMXVOUT for the probe channel multiplexers)
Back End Processor (48 VDC)
Operator Panel, LCD, XYZ motors (48 VDC)
5 - 66
2
EVISION
72
2 x 115 VAC
35,17(56
+48 VDC
TO
BEP POWER SUPPLY
+24 VDC
+/- 15 VDC
+11 VDC
+/-6V VDC
Section 5-10 - Power distribution
0$,1 32:(5 6833/<
32:(5
)$&725
&255(&7,21
,62/$7,21
$1'
9$&
'& +($'
,19(57(5
&21752/
'&'&
DC VOLTGES -> FEP'S MOTHERBOARD -> FEP'S BOARDS
&219(57(56
TX POWER TO FEP MOTHERBOARD TO RELAY CARD
' 02725
&175/
+%5,'*(
VIVID E9 / VIVID E7 BT'13 S
M
ERVICE
ANUAL

Hide quick links: