3-5-4
Connections on the I/O Rear Panel
The Peripheral/accessory connector panel (I/O Rear Panel) is located at the rear of the Ultrasound
system behind the rear door.
NOTE:
Accessory equipment connected to the analog and digital interfaces must be certified according
to the respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-
1 for medical equipment). Furthermore, all complete configurations shall comply with the valid
version of the system standard IEC60601-1-1. Everybody who connects additional equipment
to the signal input part or signal output part of VIVID E9 / VIVID E7, configures a medical
system, and is therefore responsible that the system complies with the requirements of the valid
version of IEC60601-1-1. If in doubt, consult the technical service department or your local
representative for GE.
3-5-4-1
Connect Ethernet
Connect the network cable to the Ethernet connector on the BEP I/O board.
The connector is located on the connector panel on the rear side of VIVID E9 / VIVID E7.
Figure 3-3 Ethernet connection on rear side of VIVID E9 / VIVID E7
3-5-4-2
Connect USB Flash Card
NOTE:
USB Flash Cards approved for VIVID E9 / VIVID E7 are verified for EMC performance
according to EN55011 class B. The use of any other USB Flash Cards will compromise this
verification, and may cause interference on VIVID E9 / VIVID E7 itself, or on other electronic
devices.
For approved models, see:
Install the USB Flash Card in one of the USB ports on the VIVID E9 / VIVID E7. Two ports are provided
on the Operator Panel, to the left for the Touch Screen.
2
EVISION
L
R
Section 9-15 "Peripherals for VIVID E9 / VIVID E7" on page
Chapter 3 - System setup
VIVID E9 / VIVID E7 BT'13 S
!
ETHERNET
CONNECTOR
M
ERVICE
ANUAL
9-51.