Probe Connection

Vivid 7 / Vivid 7 PRO has four positions to plug in probes. The connector to the left is a "dummy"
connector, only used for parking one probe. The three other connectors are selectable and used for
scanning.
A separate "Pedof" Doppler probe connector is also available.
Probe Connectors
Parking Slot
for probe
connector
3-7-9-1
Connect a Probe
NOTE:
It is not necessary to turn OFF Power to connect or disconnect a probe.
Do not allow the probe head to hang freely. Excessive impact to the probe will result in
CAUTION
irreparable damage.
To prevent probe connector pins, or PCB board damage, do not use excessive force when
CAUTION
connecting the probes.
1.) Hold the probe connector vertically with the cable pointing upward.
2.) Turn the connector locking handle counter-clockwise to the horizontal position.
3.) Align the connector with the probe port and carefully push into place.
4.) Rotate the locking handle to the full vertical position to lock in place. (See
5.) Position the probe cable so that it is not resting on the floor.
Take the following precautions with the probe cables:
CAUTION
Keep away from the wheels
Do not bend
Do not cross cables between probes.
3-7-9-2
Disconnect Probe
1.) Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector.
(See
2.) Remove the connector from the port.
3.) Ensure that the probe head is clean before placing the probe in its storage case, see
4 on page 10-7
02
EVISION
Figure 3-10 Probe Connectors
Figure
3-10.)
for cleaning instructions.
Chapter 3 - Installation
V
7 / V
IVID
"Pedof" Doppler Probe Connector
7 PRO S
M
IVID
ERVICE
ANUAL
Locked position
Unlocked position
Figure
3-10.)
section 10-5-
3 - 13