Input Power - GE VIVID E9 Service Manual

Hide thumbs Also See for VIVID E9:
Table of Contents
GE H
EALTHCARE
D
GA091568, R
IRECTION
10-5-2-2
Peripheral/option checks
If any peripherals or options are not part of the VIVID E9 configuration, the check can be omitted.
Refer to the User Manual for a list of approved peripherals/options.
Table 10-4
Step
1.
Media
2.
B/W Printer
3.
Color Printer
4.
DICOM
5.
ECG
6.
Footswitch
10-5-3

Input Power

10-5-3-1
Mains cable inspection
Table 10-5
Step
1.
Unplug Cord
2.
Inspect
3.
Verify
4.
Verify
10 - 8
5
EVISION
GE Approved Peripheral/Hardware Option Functional Checks
Item
Verify media drive(s) read/write properlty. Clean if necessary.
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
Verify that DICOM is functioning properly. Send an image to a DICOM device.
Verify basic operation with customer
Verify that the footswitch is functioning as programed.
Clean as necessary.
Mains Cable Inspection
Item
Disconnect the mains cable from the wall and the VIVID E9.
Inspect it and its connectors for damage of any kinds.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
Inlet connector retainer is functional.
Section 10-5 - System maintenance
VIVID E9 S
Description
Description
M
ERVICE
ANUAL
Table of Contents