Dc Offset Calibration (Front End Alignment); Introduction; When To Do A Front End Alignment; Front End Alignment Procedure - GE VIVID E9 Service Manual

Hide thumbs Also See for VIVID E9:
Table of Contents
GE H
EALTHCARE
D
GA091568, R
IRECTION
Section 6-6

DC Offset Calibration (Front End Alignment)

6-6-1

Introduction

DC Offset Calibration (Front End Alignment) is performed to calibrate each ADC channel of the GRX
inputs. The system calibrates the ADC to zero output when there is no signal on the input. These bias
voltages are stored in the Back End Processor.
6-6-2

When to do a Front End Alignment

Do the Front End Alignment:
when installing a new VIVID E9
after software has been updated or replaced
when DRX boards have been interchanged
when a DRX board has been replaced
when the Back End Processor has been replaced
6-6-3

Front End Alignment Procedure

The Front End Alignment is available on the Common Service Desktop interface.
1.) Disconnect all probes.
2.) Power on the VIVID E9.
3.) Select the InSite ExC icon (GE phone icon) on the screen to display the InSite ExC menu.
Figure 6-10 InSite ExC icon in the status bar
4.) Select Service Desktop.
Figure 6-11 InSite ExC Menu
5.) Select the Diagnostics tab.
6.) Open the Analog Receive folder.
7.) Select Execute to run the calibration.
8.) After completing the calibration, restart the VIVID E9.
5
EVISION
Chapter 6 - Service adjustments
VIVID E9 S
M
ERVICE
ANUAL