Probe/Connectors Check - GE VIVID E9 Service Manual

Hide thumbs Also See for VIVID E9:
Table of Contents
GE H
EALTHCARE
D
GA091568, R
IRECTION
4-3-7

Probe/Connectors Check

TAKE THE FOLLOWING PRECAUTIONS WITH THE PROBE CABLES:
CAUTION
- KEEP AWAY FROM THE WHEELS
- DO NOT BEND
- DO NOT CROSS CABLES BETWEEN PROBES
Table 4-5
Step
1.
2.
3.
4.
5.
6.
7.
8.
4 - 44
5
EVISION
Probe and Connectors Checks
Task
Press Probe on the Operator Panel.
If not already selected. Use the trackball to select the
desired probe.
Trackball to the desired application
Press Select to launch the application.
To change application without changing the current
probe, press Appl. on the Operator Panel.
Verify no missing channels
Verify there's no EMI/RFI or artifacts specific to the
probe.
Test the probe in each active connector slot.,
see
3-7-6 "Connecting Probes" on page 3-24
Do a leakage test on the probe, see
"Electrical Safety Tests" on page 10-20
Repeat this procedure for all available probes.
Section 4-3 - Functional checks
A list of the connected probes will pop up on the screen.
An application menu for the desired probe is listed on
the screen.
The selected application starts.
All channels are functioning.
No EMI/RFI or artifacts.
It will display pictorial data each time
Section 10-6
It passes the test.
VIVID E9 S
ERVICE
Expected Result(s)
M
ANUAL
Table of Contents
loading

Table of Contents