GE VIVID E9 Service Manual page 513

Hide thumbs Also See for VIVID E9:
Table of Contents
GE H
EALTHCARE
D
GA091568, R
IRECTION
8-7-2-2
XY-Brake Assy removal procedure (cont'd)
Figure 8-185 XY Brake Removal
8.) Gently, but firmly, pull down on the long end of the allen wrench. The brake assembly will slide out
of the slot it mates with in the brake shoe.
NOTE:
Figure 8-186 shows the XY brake removed and demonstrates placement of the Allen wrench.
Figure 8-186 XY Brake Allen wrench placement - reference