Introduction
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
1-10
When the top console is in its locked position, the gas shock is
compressed and stores mechanical energy. During normal
operation the top console, the weight of the monitor and the
mechanical force of the gas shock are in balance. Take care if/
when you activate this gas shock. Personal injury can occur
after the panel is removed and the shock pressure is released.
Take care when you repair the elevation assembly.
Risk of electrical shock, Ultrasound system must be turned off
and disconnected from power source. Cord must be controlled
at all times.
Wait for at least 20 seconds for capacitors to discharge as
there are no test points to verify isolation. The amber light on
the OP panel on/off button will turn off.
Ultrasound System components may be energized. Always
refer to the Ultrasound system's Proprietary Service Manual for
LOTO warnings and cautions.
Explosion Warning
DO NOT operate the equipment in an explosive atmosphere.
Operation of any electrical equipment in such an environment
constitutes a definite safety hazard.
Use all Personal Protection Equipment (PPE) such as gloves,
safety shoes, safety glasses, and kneeling pads, to reduce the
risk of injury.
Beware of possible sharp edges on all mechanical parts. If
sharp edges are encountered, the appropriate PPE should be
used to reduce the risk of injury.
Wear all PPE including gloves as indicated in the chemical
MSDS.
Vivid E80/E90/E95
GC091052 Rev. 3
Service Manual