GE Vivid E80 Service Manual page 139

Hide thumbs Also See for Vivid E80:
Table of Contents
Preparations
Using M-Mode
Conventional M-Mode
Anatomical M-Mode
NOTE:
Vivid E80/E90/E95
GC091052 Rev. 3
This unit has three types of M-Mode:
Conventional M-Mode (MM): displays a distance/time plot of
a cursor line in the axial plane of the 2D-image.
Anatomical M-Mode (AMM): displays a distance/time plot
from a cursor line, which is independent from the axial
plane. AMM is available in grayscale, color, TVI, Tissue
Tracking, Strain rate and Strain modes.
Curved Anatomical M-Mode (CAMM): displays a distance/
time plot from a free-drawn cursor line. CAMM is available in
grayscale, color, TVI, Tissue Tracking, Strain rate and Strain
modes.
Conventional M-Mode can be combined with Color Mode.
1. Connect one of the probes, to the Ultrasound system's
left-most probe connector.
2. Turn ON the Ultrasound system.
The 2D Mode window is displayed (default mode).
3. Press MM on the Operating Panel to bring up an M-Mode
picture on the screen.
4. Use the trackball to position the cursor over the required
area of the image.
1. To access M-Mode from any other scan mode, press MM on
the control panel.
2. Use the trackball to position the cursor over the required
area of the image.
3. Press Freeze.
4. Use the trackball to scroll through the data acquired.
1. In M-Mode or 2D-Mode Freeze, press AMM on the Touch
panel.
Anatomical M-Mode can also be used with previously
acquired digitally stored 2D images. More than one heart
cycle should be stored if performing M-Mode in post
processing.
2. Use the trackball (assigned function: Pos) to position the
cursor over the required area of the image.
Service Manual
Functional checks
4-15