Replacement Procedures
Warnings (continued)
WARNING
Preparations
8-98
Before you dispose of the hard drive, or return the BEP to the
local parts organization, make sure you remove ALL PATIENT
DATA from the hard drive, given that the hard drive is still
functional. In some countries, you may be required to delete all
software from the disk before returning the hard drive to the
parts warehouse. Follow your local policies.
When preparing for the replacement, you must perform the
following steps:
1. Record the Ultrasound system's TCPIP settings and
installed Option strings.
2. Export the Patient Archive images to a server or to external
media.
3. Back up the Report Archive, User-Defined Configuration
(Customer Presets), and Service settings.
You will perform a Restore after you have installed the new
BEP and loaded the correct software.
4. Collect the error logs and record information on why the
BEP must be changed.
5. Reload the system software from removable media. Select
No when asked: Keep User Data and Patient Archive?
6. Power down the Ultrasound system.
7. Disconnect the Mains Power Cable from the wall outlet.
8. Disconnect all probes.
9. Remove the Side Covers.
10. Remove the Top Cover.
11. Remove the B/W Printer.
12. For easier access in the next steps, remove the Front Cover.
Vivid E80/E90/E95
GC091052 Rev. 3
Service Manual