GE Vivid E80 Service Manual page 314

Hide thumbs Also See for Vivid E80:
Table of Contents
Replacement Procedures
WARNING
Preparations
Remove the GTX board
Install the GTX board
8-92
Prior to removing the Card Rack boards, the LEDs on the end
of each board should be unlit, to indicate boards are
unpowered.
When preparing for the replacement, you must perform the
following steps:
1. Power down the Ultrasound system.
2. Disconnect the Mains Power Cable from the wall outlet.
3. Disconnect all probes and external I/O cabling.
4. Remove the Right Side Cover.
5. Remove the FEP Cover.
6. Remove the Front Plane boards.
Follow these steps to remove the GTX board:
1. Pull out the GTX Board.
2. Place it on an ESD safe place.
Follow these steps to install the GTX Board:
1. Carefully align the GTX Board with the rails and push it in till
it is seated in the Back Plane connectors.
2. Carefully install the two Front Plane Boards.
3. Install the FEP Cover.
4. Install the Right Side Cover.