Table 4-10: Hart Failsafe And Write Protect Jumpers; Table 4-11: Foundation Fieldbus Simulation And Write Protect Jumpers - Honeywell SLG 700 User Manual

Smartline level transmitter guided wave radar
Hide thumbs Also See for SLG 700:
Table of Contents
Arrangements

Table 4-11: FOUNDATION Fieldbus Simulation and Write Protect Jumpers

Changes to configurations and calibrations are disabled when the
Note:
jumpers are set in the write-protect positions. Write-protect includes
disabling password access.
Page 110

Table 4-10: HART Failsafe and Write Protect Jumpers

Jumper
Failsafe = UP (High)
Write Protect = OFF (Not Protected)
Failsafe = DOWN (Low)
Write Protect = OFF (Not Protected)
Failsafe = UP (High)
Write Protect = ON (Protected)
Failsafe = DOWN (Low)
Write Protect = On (Protected)
Image
Fieldbus Simulation Mode = OFF
Write Protect = OFF (Not Protected)
Fieldbus Simulation Mode = OFF
Write Protect = ON (Protected)
Fieldbus SIM Mode = ON
Write Protect = OFF (Not Protected)
SLG 700 SmartLine Level Transmitter User's Manual
Description
Description
Revision 7
Table of Contents
loading