Trimming The Analog Output - Emerson Rosemount 2051 Reference Manual

Pressure transmitter with hart revision 5 and 7 selectable protocol
Hide thumbs Also See for Rosemount 2051:
Table of Contents
Operation and maintenance
February 2024
Recall factory trim using a communication device
Procedure
1. From the HOME screen, enter the fast key sequence:
2. Follow the steps within the communication device to complete the sensor trim.
Recall factory trim using AMS Device Manager
Procedure
1. Right-click the device and go to Method → Calibrate → Restore Factory Calibration.
2. Set the control loop to Manual.
3. Select Next.
4. Select Sensor Trim under Trim to recall and click Next.
5. Follow the screen prompts to recall sensor trim.
Recall factory trim using a local operator interface (LOI)
Refer to
Figure 5-4: Recall factory trim using LOI
VIEW CONFIG
ZERO TRIM
UNITS
RERANGE
LOOP TEST
DISPLAY
EXTENDED MENU
EXIT MENU
5.7 

Trimming the analog output

You can use the
output at the 4 and 20 mA (1 – 5 Vdc) points to match the plant standards.
Perform this trim after the digital to analog conversion so only the 4–20 mA analog (1– 5
Vdc) signal will be affected.
curve is affected when an analog output trim is performed.
90
3, 4, 3
Fast keys
Figure 5-4
to recall factory sensor trim.
EXTENDED MENU
CALIBRAT
DAMPING
TRANSFER FUNCT
SCALED VARIAB
ASSIGN PV
TAG
ALARM SAT VALUES
PASSWORD
SIMULATE
HART REV
BACK TO MENU
EXIT MENU
Analog Output Trim
Figure 5-5
CALIBRAT
ZERO TRIM
LOWER TRIM
UPPER TRIM
ANALOG TRIM
FACTORY RECALL
BACK TO MENU
EXIT MENU
command to adjust the transmitter's current
graphically shows the two ways the characterization
Reference Manual
00809-0100-4107
FACTORY RECALL
SENSOR RECALL
ANALOG RECALL
BACK TO MENU
EXIT MENU
Rosemount 2051