4

Product certifications

Rev 3.2
4.1
European Directive Information
A copy of the EU Declaration of Conformity can be found at the end of the
Quick Start Guide. The most recent revision of the EC Declaration of
Conformity can be found at Emerson.com/Rosemount.
4.2
Ordinary Location Certification from FM Approvals
As standard, the transmitter has been examined and tested to determine
that the design meets the basic electrical, mechanical, and fire protection
requirements by a nationally recognized test laboratory (NRTL) as accredited
by the Federal Occupational Safety and Health Administration (OSHA).
4.3
North America
E5 FM Explosionproof and Dust Ignitionproof
Certificate
Standards
used
Markings
I5 FM Intrinsic Safety and Nonincendive
Certificate
Standards
used
Markings
Special Conditions for Safe Use(X):
1. The Model 3051 transmitter housing contains aluminum and is
considered a potential risk of ignition by impact or friction. Care must
be taken into account during installation and use to prevent impact
and friction.
Quick Start Guide
0T2H0.AE
FM Class 3600 – 2011, FM Class 3615 – 2006, FM Class
3616 – 2011, FM Class 3810 – 2005, ANSI/NEMA 250 –
2008
XP CL I, DIV 1, GP B, C, D; DIP CL II, DIV 1, GP E, F, G; CL III;
T5(-40 °C ≤ Ta ≤ +85 °C); Factory Sealed; Type 4X
1Q4A4.AX
FM Class 3600 – 2011, FM Class 3610 – 2010, FM Class
3611 – 2004, FM Class 3810 – 2005, ANSI/NEMA 250 –
2008
IS CL I, DIV 1, GP A, B, C, D; CL II, DIV 1, GP E, F, G; Class III;
DIV 1 when connected per Rosemount drawing
03031-1019; NI CL 1, DIV 2, GP A, B, C, D; Temp code T4;
Type 4x; Factory sealed
Quick Start Guide
19