Connecting The Unit - GE Vivid i User Manual

Hide thumbs Also See for Vivid i:
Table of Contents
Getting started
Electromagnetic interferences
The Vivid i ultra-
Ensure that the unit is protected from electromagnetic
sound unit is ap-
interferences as follows:
proved for use in
hospitals, clinics
and other environ-
mentally qualified
facilities, in terms of
the prevention of ra-
dio wave interfer-

Connecting the unit

ence. Operation of
the unit in an inap-
A GE-qualified person should perform the initial system
propriate environ-
ment can cause
installation.
electronic interfer-
Connecting the Vivid i ultrasound unit involves preliminary
ence to radios and
television sets situ-
checks of the power adaptor unit and cord, voltage level and
ated near the medi-
compliance with electrical safety requirements.
cal equipment.
Use only power supply cords, cables and plugs provided by or
designated by GE Medical Systems.
Ensure that the power cord and plug are intact and that the
power plug is the proper hospital-grade type (where required).
When using the mains outlets, the unit should be connected to
a fixed power socket which has the protective grounding
connector. Never use an extension cord or adapter plug.
Failure to provide an adequate earth circuit can cause electrical
shock, resulting in serious injury.
WARNING
10
Operate the unit at least 4.5 meters (fifteen feet) away from
equipment that emits strong electromagnetic radiation.
Shield the unit when operating it in the vicinity of radio
broadcasting equipment, if necessary.