GE Medical Systems
REV1
2–4
Special Material Specifications
SECTION 3
ENVIRONMENT
TABLE 4–4
AMBIENT CONDITIONS
Composition of materials in contact with the breast.
Transparent plastics used in compression paddles are of polycarbonate Lexan.
Bucky plates are of carbon fiber or plastic material.
HUMIDITY
In use
(in original crates)
Min.
Max.
Min.
30%
75%
10%
TEMPERATURE
In use
(in original crates)
Min.
Max.
Min.
10° C
40° C
20° C
50°F
104°F
68°F
ALTITUDE
In use
Storage
+ 3000 m
+ 3000 m
9842'
9842'
4-10
Storage
Max
95%
Storage
Max.
70° C
158°F
SENOGRAPHE DMR+
om 5128689-2-100