GE SENOGRAPHE DMR+ Technical Publications
Hide thumbs Also See for SENOGRAPHE DMR+:
Technical
Publications
0459
5128689–2–100
Revision 1
SENOGRAPHE DMR+
om
Operator Manual
do not duplicate
E
Copyright
2005 GE Medical Systems
Table of Contents
loading

Summary of Contents for GE SENOGRAPHE DMR+

  • Page 1 Technical Publications 0459 5128689–2–100 Revision 1 SENOGRAPHE DMR+ Operator Manual do not duplicate Copyright 2005 GE Medical Systems...
  • Page 2 WARNING X–RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS MEASURES OF PROTECTION ARE STRICTLY OBSERVED Though this equipment is built to the highest standards of electrical and mechanical safety, the useful x–ray beam becomes a source of danger in the hands of the unauthorized or unqualified operator. Excessive exposure to x–radiation causes damage to human tissue.
  • Page 3 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Services Offices of GE Medical Systems For service issues, contact the Service Office for your country. AMERICAS UNITED STATES OF AMERICA GE Medical Systems Headquarters Tel: +1–414–544–3011 3000 North Grandview Blvd. WAUKESHA, WI 53188 Mail:P.O.
  • Page 4 GE Medical Systems Danmark Tel: +45 45 51 00 55 Skovlytoften 4 Fax: +45 42 42 59 89 DK–2840 HOLTE DEUTSCHLAND (Germany) GE Medical Systems Deutschland GmbH & Tel: 49–6102–36–0 Co. KG Fax: +49 610–36–2588 Martin–Behaim–Strasse 10 D–63263 NEU_ISENBURGS ESPAÑA (Spain) GE Medical Systems España...
  • Page 5 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 PORTUGAL (Portugal) GE Medical Systems Portuguesa S.A. Rua Sà da Bandeira, 585 Tel: +351 2 2007696/97 Apartado 4094 Fax: +351 2 2084494 P–4002 PORTO CODEX Tlx: 22804 ROSSIYA (Russia) GE Medical Systems Tel:+7 095 935 72 41 Kosmodamianskaya nab.
  • Page 6 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Blank page.
  • Page 7 Directive. For a system, the location of the CE marking label is described in the system manual. European registered place of business: GE Medical Systems Europe Quality Assurance Manager BP 34 F 78533 BUC CEDEX France...
  • Page 8 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ELECTROMAGNETIC COMPATIBILITY (EMC) This equipment complies with IEC60601–1–2 Edition 2 EMC standard for medical devices. This equipment generates, uses, and can radiate radio frequency energy. The equipment may cause radio frequency interference to other medical and non-medical devices and radio communications.
  • Page 9 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 RECYCLING: Machines or accessories at end–of–life: The elimination of machines and accessories must be in accordance with national regulations for waste processing. All materials and components that could pose a risk to the environment must be removed from the end–of–life machines and accessories (examples: dry and wet cell batteries,...
  • Page 10 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Blank page viii...
  • Page 11 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 TABLE OF CONTENTS CHAPTER TITLE PAGE REVISION HISTORY ............
  • Page 12 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 TABLE OF CONTENTS CHAPTER TITLE PAGE MAINTENANCE .............
  • Page 13 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 TABLE OF CONTENTS (CONT.) CHAPTER TITLE PAGE 4–7 Exposure Interlock Disable ..........
  • Page 14 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Blank page...
  • Page 15 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 FOREWORD This manual is provided for SENOGRAPHE DMR operators. It is designed to supply all the information required for the correct use of this equipment. xiii...
  • Page 16 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Blank page...
  • Page 17 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Left in English intentionally REVISION HISTORY DATE TYPE OF MODIFICATION June 2005 New release based on document 5128689-100 rev. 1, with the following changes: WEEE Compliance: Chapter 4 Specifications, added Section 4 Meaning of Symbols, including symbol and text for waste of electrical and electronic equipment.
  • Page 18 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 LIST OF EFFECTIVE PAGES PAGE REVISION PAGE REVISION PAGE REVISION Title Page 2-1 thru 2-8 5-1 thru 5-2 i thru xvi 3-1 thru 3-28 1-1 thru 1-34 4-1 thru 4-12 This document is using the Conditional selection. It is a slave document.
  • Page 19 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 CHAPTER 1 - INTRODUCTION SECTION 1 PRESENTATION The SENOGRAPHE DMR is an x-ray system used primarily for mammography examinations in the standing or sitting position and under various views and magnification factors.
  • Page 20 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-1 THE SENOGRAPHE DMR...
  • Page 21 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 SECTION 2 DESCRIPTION See Illustration 1–1. The SENOGRAPHE DMR is composed of the following components: X ray system. Control console. Power supply Cabinet. Protective screen. Note: The CE marking label is located on the bottom left hand side of the power supply...
  • Page 22 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-2 X RAY SYSTEM...
  • Page 23 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–1 X-ray System See Illustration 1–2. The x-ray system is equipped with a Telescopic Column which supports the Examination Arm. The Telescopic Column must be secured to the floor using the fixed baseplate.
  • Page 24 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-3 TUBE SUPPORT ARM...
  • Page 25 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–1–1 Tube–Support Arm Consists of: X-ray Tube and Tube Housing Assembly 2 which has an Up/Down (↑↓) control located at the top of the Assembly. The Collimator 1 consists of the diaphragm and light centering controls.
  • Page 26 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-4 RECEPTOR SUPPORT ARM...
  • Page 27 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–1–2 Receptor-Support Arm See Illustration 1–4. The Receptor-Support Arm consists of the following: Guide Rail 3 and Compression system including variable SID, magnification plate, and compression paddle. SID change is made via one of the two levers 5 located at the bottom of the handrails, one on either side of the column.
  • Page 28 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-4 RECEPTOR SUPPORT ARM 1-10...
  • Page 29 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Note: The top holes are exclusively reserved for the installation of Stereotix 2 or Senovision Stereotactic positioner, or for making exposures of specimens or core biopsy samples. When making exposures of specimens or core biopsy samples, the magnification plate can be snapped onto the top holes with the Positioning Arm remaining at 0 degree.
  • Page 30 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-5 FILM HOLDER ASSEMBLY 1 2 3 1-12...
  • Page 31 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–1–3 Film-Holder Assembly See Illustration 1–5. The Film-Holder Assembly 5 can receive both the 18x24 and 24x30 Bucky and the cassette holder for magnification and is equipped with the photocell of the AEC.
  • Page 32 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-5 FILM HOLDER ASSEMBLY 1 2 3 1-14...
  • Page 33 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Front adjustment of the photocell position: Is accomplished via a level 4 located on either side of the Film Holder Assembly. Five positions are available to suit patient anatomy. Note: An Up/Down button is located on the cover of the x-ray Tube/Housing Assembly.
  • Page 34 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-6 POWER SUPPLY CABINET 1-16...
  • Page 35 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–2 Power Supply Cabinet The main contents of the Power Supply Cabinet are the generator and the system management electronics. The Cabinet consists of the following components: Cabinet box Casters. Cable raceways.
  • Page 36 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–3 Accessories The accessories delivered with the basic configuration of the SENOGRAPHE DMR consist of: 18 x 24 cm Bucky Cassette holder for magnification Square spot compression paddle 18 x 24 cm compression paddle...
  • Page 37 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–4 Optional Accessories The optional accessories for the SENOGRAPHE DMR consist of (not limited to): 24 x 30 cm Bucky. 24 x 30 cm compression paddle. STEREOTIX 2, or SENOVISION (Digital Spot & Digital Stereotaxy).
  • Page 38 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-7 CONTROL CONSOLE Control Console CDRH GE Medical Systems Control Console 2001 (in forward production from february 2002) 1-20...
  • Page 39 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–5 Control Console See Illustration 1–7 The Control Console is the interface between the Operator and the SENOGRAPHE DMR. It provides two functions: Receives operator commands. Displays replies and/or machine messages. The Control Console consists of: The Control Console, which can be placed on a counter, on the Power Supply Cabinet, or installed on the protective lead screen.
  • Page 40 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-8 CONTROL KEYBOARD Control Console CDRH This symbol may be replaced by Control Console 2001 (in forward production from february 2002) 1-22...
  • Page 41 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–5–1 Control Keyboard See Illustration 1–8 The Control Keyboard consists of the following: SENOGRAPHE DMR `ON' SENOGRAPHE DMR `OFF' SET UP Menu FLUO readout Focal Spot selection Focal Track selection Filter selection...
  • Page 42 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-9 LEFTHAND SECTION OF CONTROL KEYBOARD Control Console CDRH This symbol may be replaced by Control Console 2001 (in forward production from february 2002) 1-24...
  • Page 43 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 SECTION 3 CONTROL KEYBOARD 3–1 Lefthand Section of Control Keyboard See Illustration 1–9 The lefthand section of the Control Console Keyboard is used to initialize or select generator parameters. SENOGRAPHE DMR `ON' The indicator lamp is illuminated.
  • Page 44 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Focal Spot selection Small and large focal spot selection. Press key to toggle between small or large focal spot (0.1 or 0.3, respectively). The readout indicates the size of the selected focal spot in the language programmed at installation.
  • Page 45 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Film Density Control (FDC) decrements In AOP and AEC modes, press key to reduce density correction by one step. Film Density Control (FDC) increments FDC+ In AOP and AEC modes, press key to increase density correction by one step.
  • Page 46 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-10 CENTRAL SECTION OF CONTROL KEYBOARD Control Console CDRH Control Console 2001 (in forward production from february 2002) 1-28...
  • Page 47 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 3–2 Central Section of Control Console Keyboard The central section of the Control Console Keyboard is used to select and display the kV and mAs values. kV and mAs Readout In AOP mode the readout displays the kV values selected by the system and the mAs values output.
  • Page 48 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-10 CENTRAL SECTION OF CONTROL KEYBOARD (COND.) Control Console CDRH Control Console 2001 (in forward production from february 2002) 1-30...
  • Page 49 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 AOP, AEC, and Manual Mode selection The key selects the AOP, AEC and manual modes successively. The AOP mode is broken down into three positions, allowing three priorities: CNT: contrast priority. STD: standard (dose/contrast compromise).
  • Page 50 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 1-11 RIGHTHAND SECTION OF CONTROL KEYBOARD Control Console CDRH Control Console 2001 (in forward production from february 2002) 1-32...
  • Page 51 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 3–3 Righthand Section of Control Keyboard See Illustration 1–11. This section is used to prepare and initiate an exposure, control the centering light, and for decompression and system reset. Exposure Enable lamp When illuminated, exposure is enabled.
  • Page 52 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Exposure button Press and hold to make exposure. To stop exposure immediately, release key. A buzzer sounds at the end of exposure. Note: In AOP mode and following preparation (anode rotation), exposure is made in two...
  • Page 53 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 CHAPTER 2 - PREPARING AN EXAMINATION SECTION 1 PREPARING THE EQUIPMENT Set the image receptor, focal distance, field size, and compression paddle for the required view. For standard exposures, use 18x24 cm or 24x30 cm cassettes and a compression paddle suitable to the format selected.
  • Page 54 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 TABLE 2–1 TRACK/FILTER CONFIGURATIONS TRACK FILTER KV RANGE 22 thru 35 22 thru 40 22 thru 49 25 thru 49 25 thru 49 As image contrast depends on kV values used, note that while an increase in the kV value may reduce the radiation dose to the patient, it will reduce the contrast of the image.
  • Page 55 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Compression It is advised that sufficient breast compression be used to benefit the following image quality/dose reduction advantages: Compression reduces motion blurring. Compression reduces geometric unsharpness by ensuring direct contact between breast and image receptor, and by spreading apart glandular breast tissue.
  • Page 56 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 SECTION 3 EXPOSURE MODE 3–1 AOP Mode The AOP (Automatic Optimization Parameters) Mode controls radiation and allows the system to select the main parameters. It optimizes the track/filter/kV configuration as a function of the required image quality, and of the thickness and density of the breast being examined.
  • Page 57 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Note: In magnification, use of the bucky is not recommended. The operator should use the cassette holder (refer to Chap 1, Para 2–1–3). Note: Always make sure that the magnification plate is correctly locked in place before starting an exam.
  • Page 58 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 3–2–2 Magnification Note: Use of the bucky is not recommended in magnification. The operator should use the cassette holder. Note: Always make sure that the magnification plate is correctly locked in place before starting an exam.
  • Page 59 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 The operator selects the following parameters: Focal track (Mo, Rh) Filter (Mo, Rh, A1) Focal spot (0.3 or 0.1) Screen–film combination, kV values, mAs values. The system displays the inhibit conditions. Note: Selecting the screen-film combination in Manual Mode has no effect.
  • Page 60 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 3–3–2 Magnification Note: Use of the Bucky is not recommended in magnification. The operator should use the cassette holder. Note: Always make sure that the magnification plate is correctly locked in place before starting an exam.
  • Page 61 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 CHAPTER 3 - MAINTENANCE SECTION 1 CLEANING & DISINFECTION GENERAL INFORMATION Adequate cleaning and disinfection is necessary to prevent disease transmission. Be CAUTION sure to thoroughly clean and disinfect equipment surfaces that contact the patient and all equipment surfaces likely to become soiled during use.
  • Page 62 RECOMMENDED GERMICIDES The following legally marketed products have been used on GE equipment without causing equipment damage. Low or Intermediate Disinfection , manufactured by Calgon Vestal Laboratories, St. Louis, MO, U.S.A., EPA Reg. No.
  • Page 63 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 High Level Disinfection CIDEX, manufactured by Johnson & Johnson Medical, Inc., Arlington, TX, U.S.A., EPA Reg. No. 7078–1, EPA Est. No. 36126–PR–1 (510(k) K924434). CIDEX CONTAINS GLUTARALDEHYDE. DIRECT CONTACT DANGER IS CORROSIVE TO EXPOSED TISSUE, CAUSING EYE DAMAGE AND SKIN IRRITATION DAMAGE.
  • Page 64 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 SECTION 2 PREVENTIVE MAINTENANCE Despite its inherent qualities, the Senographe “DMR” requires minimum maintenance. A maintenance program, whose frequency varies according to the type of use of the equipment, is outlined in Table 3–1.
  • Page 65 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 SECTION 3 MESSAGES Three types of messages may be displayed for the user on the gantry or on the Control Console: I xx: Information message: fault message which does not inhibit the exposure.
  • Page 66 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 TABLE 3–2 ERROR MESSAGES DISPLAYED ON READOUT AND ON CONTROL CONSOLE Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message I 07 None SELECTED FILTER NOT Filter selected by operator...
  • Page 67 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message I 29 None EXCESSIVE mAs FOR The amount of mAs Switch to a different mode or to THIS MODE xx mAs necessary to repeat the a different filter if in AEC mode.
  • Page 68 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message I 72 COLUMN FAIL COLUMN MOTION Elevator is not functioning -Switch OFF, and switch ON. FAILURE correctly .
  • Page 69 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message I 86 DECOMPRESS DECOMPRESSION Compression paddle hit an 1. Raise paddle to park using obstacle during decompression pedal.
  • Page 70 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message E 03 None ROTOR FAILURE Failure in anode starter. -Switch OFF, and switch ON. If fault continues, note ALL messages and contact GEMS Field Service.
  • Page 71 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message E 12 None CHECKSUM ERROR Calibration parameters have 1. Switch OFF. (GEN) been modified. 2. Switch ON. 3. If fault continues, note ALL messages and inform GEMS Field Service.
  • Page 72 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message E 58 COLLIM FAIL FRONT BLADE FAILURE Front collimator blade is 1. Switch OFF. faulty, so a good format 2.
  • Page 73 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message S 08 None REQUIRED POWER IS Power requested in Manual - Lower kV or mAs values. TOO HIGH Mode is too high.
  • Page 74 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Message Message on Message on Explanation/meaning Corrective Action Column Readout Control Console of message S 32 None COMPRESSION Breast thickness is badly 1. Switch OFF. DETECTOR FAULT calculated in AOP mode. 2. Switch ON.
  • Page 75 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 SECTION 4 MEDICAL PROGRAMMING 4–1 GENE (language) 4–1–1 Access to Programming Mode To access the programming mode, press The APPLICATION Menu, focal spot, focal track, filter, etc., disappears Press, and key lights up...
  • Page 76 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 4–1–3 Confirm, & Return to APPLICATION Menu To confirm the function selected and return to the APPLICATION Menu (daily use), press repeatedly until focal spot, focal track, filter, etc” return. Press repeatedly until daily use display...
  • Page 77 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 4–2 Arm: Speed of Compression 4–2–1 Access to Programming Mode To access the programming mode, press The APPLICATION Menu, focal spot, focal track, filter, etc., disappears. Press, and key lights up APPLICATION SETUP 4–2–2...
  • Page 78 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 APPLICATION SETUP COMPRESSION SPEED: FAST FAST SLOW Press to display speed selected 4–2–3 Confirm, & Return to APPLICATION Menu To confirm the function selected and return to the APPLICATION Menu (daily use), press repeatedly until focal spot, focal track, filter, etc”...
  • Page 79 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 4–3 Arm: Compressive Force 4–3–1 Access to Programming Mode To access the programming mode, press The APPLICATION Menu, focal spot, focal track, filter, etc., disappears. Press, and key lights up APPLICATION SETUP 4–3–2...
  • Page 80 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 APPLICATION To increment by 1 daN maximum = 20 daN. SETUP MAX COMPRESS. FORCE = 10 daN To decrement by 1 daN VALID minimum = 4 daN. To confirm. Press selected key 4–3–3...
  • Page 81 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 4–4 Arm: Automatic Decompression 4–4–1 Access to Programming Mode To access the programming mode, press The APPLICATION Menu, focal spot, focal track, filter, etc., disappears. Press, and key lights up APPLICATION SETUP 4–4–2...
  • Page 82 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 APPLICATION SETUP END OF EXPOSURE DECOMP = YES Press to display reply selected 4–4–3 Confirm, & Return to APPLICATION Menu To confirm the function selected and return to the APPLICATION Menu (daily use), press repeatedly until focal spot, focal track, filter, etc”...
  • Page 83 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 4–5 Arm: Beep 4–5–1 Access to Programming Mode To access the programming mode, press The APPLICATION Menu, focal spot, focal track, filter, etc., disappears. Press, and key lights up APPLICATION SETUP 4–5–2 Selecting a Function To select a function, press the key located under the symbol for that function.
  • Page 84 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 APPLICATION SETUP WITH BEEP: YES Press to display reply selected 4–5–3 Confirm, & Return to APPLICATION Menu To confirm the function selected and return to the APPLICATION Menu (daily use), press repeatedly until focal spot, focal track, filter, etc” return.
  • Page 85 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 4–6 Decompression Height 4–6–1 Access to Programming Mode To access the programming mode, press The APPLICATION Menu, focal spot, focal track, filter, etc., disappears. Press, and key lights up APPLICATION SETUP 4–6–2 Selecting a Function To select a function, press the key located under the symbol for that function.
  • Page 86 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 APPLICATION To increment by 1 cm maximum = 40 cm. SETUP DECOMPRESSION HEIGHT = 25 CM To decrement by 1 cm VALID minimum = 4 cm. To confirm. Press selected key 4–6–3 Confirm, &...
  • Page 87 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 4–7 Exposure Interlock Disable The Senographe DMR includes a cassette exposure interlock which prevents an exposure from being made if there is no cassette in the bucky or if the cassette used for the previous exposure has not been removed.
  • Page 88 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 APPLICATION SETUP MEDICAL SETUP SPEED FORCE DECOMP BEEP HEIGHT FILM Press to display exposure interlock selection menu * Press "YES" to disable exposure if there is APPLICATION no cassette in the bucky, or if the cassette has not been changed.
  • Page 89 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 CHAPTER 4 - SPECIFICATIONS SECTION 1 PHYSICAL SPECIFICATIONS See Illustrations 4–1, 4–2, 4–3 and 4–4. TABLE 4–1 SPACE REQUIREMENTS & WEIGHT Description Illustration Dimensions Weight Length Width Height kg (lb) mm (ft)
  • Page 90 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 4–1 GANTRY: SPACE REQUIREMENTS 180° 180° 400 1.31' 150 mini 1200 3.94' 0.50'min 0.74' VIEW A 1500 mini 4.93' min 1500 mini 4.93' min...
  • Page 91 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 4–2 CONTROL CABINET: SPACE REQUIREMENTS 1.84' 431 1.41' 295 0.97' 1.81' 0.5 m...
  • Page 92 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 4–3 PROTECTIVE SCREEN/CONTROL CONSOLE: SPACE REQUIREMENTS 28" 275 11" 80 3.20"...
  • Page 93 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 ILLUSTRATION 4–4 ACCESSORY TROLLEY: SPACE REQUIREMENTS 900 2.95' 0.5 m...
  • Page 94 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 SECTION 2 TECHNICAL SPECIFICATIONS 2–1 Electrical Specification The Senographe DMR is a class I type B equipment. 2–1–1 Line voltage specifications Single–phase input voltage: * 200/208/220/240/380/415 V (± 10 %) Line frequency specifications 50 or 60 Hz (1 Hz) 2–1–2...
  • Page 95 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 The generator can supply a maximum of 100 W to the tube. This example represents one 30 kV/200mAs exposure each minute. 2–1–6 Line Conditioner Resonance Frequency 20 kHz approx. 2–1–7 Generator Power (excluding tube) 5 kW maximum.
  • Page 96 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–2 Filters TRACK FOCAL SPOT Large 100 mA max. 75 mA max. Small 40 mA max. 40 mA max. The filters are installed on a disk driven by a stepping motor which moves from one filter to the other.
  • Page 97 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–3 Beam-limiting Devices The available x-formats are as follows: TABLE 4–2 FIELD COVERED BY BEAM Format Focal Spot Large Large 18 x 24 18 x 24 18 x 24 18x24 Large Small...
  • Page 98 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 2–4 Special Material Specifications Composition of materials in contact with the breast. Transparent plastics used in compression paddles are of polycarbonate Lexan. Bucky plates are of carbon fiber or plastic material. SECTION 3 ENVIRONMENT TABLE 4–4...
  • Page 99 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 SECTION 4 MEANING OF SYMBOLS Protective earth (ground) Earth (ground) Dangerous voltage Type B equipment This symbol indicates that waste electrical and electronic equipment must not be disposed of as unsorted munic- ipal waste and must be collected separately.
  • Page 100 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Blank page 4-12...
  • Page 101 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 CHAPTER 5 - REGULATIONS The SENOGRAPHE DMR complies to the following regulations: UL 187 X-ray equipment CSA 22.2, No. 114 Radiation Emitting Equipment IEC 60601–1 Medical Electrical Equipment Part 1: General requirements for safety IEC 60601–2–7...
  • Page 102 GE Medical Systems SENOGRAPHE DMR+ REV1 om 5128689-2-100 Blank page...