GE SENOGRAPHE DMR+ Technical Publications page 72

Hide thumbs Also See for SENOGRAPHE DMR+:
Table of Contents
GE Medical Systems
REV1
Message
Message on
No.
Column Readout
E 58
COLLIM FAIL
E 64
FILTER
E 71
None
E 86
CHKSUM ERROR
S 01
None
S 02
None
S 03
None
S 05
None
S 06
None
Message on
Control Console
FRONT BLADE FAILURE
FILTER POSITIONING
FAILURE
BUCKY FAILURE
ARM CHECKSUM
ERROR
ABORTED EXPOSURE
EXPOSURE ABORTED
BY HV ARCING
EXPOSURE ABORTED
BY AEC
WRONG SELECTION OF
HV/TRACK COUPLE
kV TOO LOW
Explanation/meaning
of message
Front collimator blade is
faulty, so a good format
cannot be ensured.
Filter rotation is out of order,
so good filter positioning is
not ensured.
Exposure is inhibited.
Bucky cannot be moved
under normal conditions.
Content of saved memory
has been modified.
Calibration may be false.
Compression is inhibited.
Exposure in progress has
been aborted because
Exposure Button was
released by operator before
the end of exposure.
Exposure was aborted by
equipment because in
excess of eight x ray tube
arcing were produced during
the exposure so far.
Exposure aborted by AEC
because not enough mAs
available to make exposure.
Selected kV/focal track is not
enabled (kV values are too
low in the Rh focal track).
kV values are too low to use
AEC (should exceed 24 kV).
3-12
SENOGRAPHE DMR+
om 5128689-2-100
Corrective Action
1. Switch OFF.
2. Switch ON.
3. If fault continues, note ALL
messages and inform
GEMS-E Field Service.
1. Switch OFF.
2. Switch ON.
3. If fault continues, note ALL
messages and inform
GEMS Field Service.
1. Remove Bucky.
2. Reinstall Bucky.
3. Make a Rad Prep.
4. If failure continues, note ALL
messages, and inform
GEMS Field Service.
Exposures can be made without
the Bucky.
1. Switch OFF.
2. Switch ON.
3. If fault continues, note ALL
messages and inform
GEMS Field Service.
1. Press RESET.
2. Start exam again.
1. Press RESET.
2. Start exam again.
1. Press RESET.
2. Change the parameters.
3. Start exam again.
-Change the configuration.
-Change kV values.