GE SENOGRAPHE DMR+ Technical Publications page 47

Hide thumbs Also See for SENOGRAPHE DMR+:
Table of Contents
GE Medical Systems
REV1
3–2
Central Section of Control Console Keyboard
The central section of the Control Console Keyboard is used to select and display the kV
and mAs values.
1
kV and mAs Readout
D
In AOP mode the readout displays the kV values selected by the system and the
mAs values output.
D
At start of compression, the kV readout displays the code of the mode selected:
CNT, DOSE, and STD. The mAs readout displays the AUTO mode.
D
At end of exposure, the kV readout displays the kV values selected. The kV
readout displays the first letter of the code of the selected mode (C, D, or S) and
the mAs readout displays the mAs values output.
D
In AEC mode, the readout displays the kV values selected by the operator and
the mAs values output preceded by the letter A.
D
In manual mode the readout displays the kV and mAs values selected by the
operator.
2
kV selection
The kV values are displayed 22 thru 49 in steps of 1 kV via the knurled knob (28
positions). At end of travel (there is no mechanical stop), the readout displays
the maximum or minimum kV values (22 kV or 49 kV).
In AOP mode, rotating the knob has no effect.
3
mAs selection
The mAs values are displayed 4 thru 600 mAs via the knurled knob. At
end of travel (there is no mechanical stop), the readout displays the minimum
or maximum mAs values (4 or 600 mAs).
In AOP and AEC mode, turning the knob has no effect.
1-29
SENOGRAPHE DMR+
om 5128689-2-100
Table of Contents
loading