Tether Handling

The tether connects the detector to the system. Damage to the tether means that the detector cannot
send image data to the unit.
To ensure maximum tether life, follow these guidelines:
Treat the tether with care.
Clean the tether regular with an approved chemical. Refer to
information.
Untwist the tether periodically. The colored stripe on the tether shows if it is twisted.
Do not jerk or pull on the tether. For example, do not pull on the tether to free it from under a wheel
(especially if the tether is twisted).
Do not step on the tether.
Do not run over the tether with the unit, cart, table or other equipment.
Do not use the tether as a handle.
Do not bend or fold the tether sharply, especially at the points where the tether connects to the
detector and to the unit.

Grid Handling

The grid is a sensitive device containing delicate mechanical structures. It is made of very thin pieces of
lead (Pb) and interspersed with Aluminum (Al).
Do not drop the grid.
Do not allow objects to hit/impact the grid.
Do not immerse grid in water or other liquid.
Do not peel labels or vinyl covering.
Inspect the grid for physical damage before use.
Clean the grid regularly (between each patient) with an approved chemical. Refer to
10) for more information.
Definium AMX 700 X-Ray System
Operator Manual
5161515-1EN
Rev. 6 (10 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter A1: Detector Overview
Cleaning
(p. A1-10) for more
Cleaning
(p. A1-
A1-9