GE Voluson 730Pro Service Manual
Hide thumbs Also See for Voluson 730Pro:
Table of Contents
Technical
Publication
Direction KTI105947
Revision 2
GE Medical Systems - Kretztechnik
Voluson® 730Pro / 730ProV (BT´04)
Service Manual
• Voluson® 730Pro / 730ProV systems with Serial number A31501 - A33999
• Voluson® 730Pro / 730ProV systems with Software version SW 4.x.x
• Voluson® 730Pro / 730ProV systems that were upgraded to BT´04
• Voluson® 730Pro / 730ProV systems, upgraded to BT´05 (SW 5.x.x)
0123
©
Copyright
2004, 2005 by General Electric Co.
GE Medical Systems
Kretz Ultrasound
Table of Contents
loading

Summary of Contents for GE Voluson 730Pro

  • Page 1 Technical Publication Direction KTI105947 Revision 2 GE Medical Systems - Kretztechnik Voluson® 730Pro / 730ProV (BT´04) Service Manual • Voluson® 730Pro / 730ProV systems with Serial number A31501 - A33999 • Voluson® 730Pro / 730ProV systems with Software version SW 4.x.x •...
  • Page 3 • VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN. Important Precautions...
  • Page 4 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
  • Page 5 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Important Precautions...
  • Page 6 GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 7 The Material is protected by Copyright and Trade Secret laws; the violation of which can result in civil damages and criminal prosecution. If you are not party to such a license agreement or a GE Medical Systems Employee, you must exit this Material now.
  • Page 8 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. Important Precautions...
  • Page 9 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change September 30, 2004 Initial Release (from Software Version 4.0.x onwards) general update and modification + implementation of service conditions August xx, 2005 for systems upgraded to BT´05 (SW 5.x.x)
  • Page 10 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. viii Revision History...
  • Page 11 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 12 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL CHAPTER 2 Pre-Installation Overview ............2 - 1 Purpose of Chapter 2 .
  • Page 13 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL CHAPTER 3 Installation Overview............3 - 1 The Purpose of Chapter 3 .
  • Page 14 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Transducer Connection ........3 - 22 Printer Installation .
  • Page 15 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of Chapter 4 .
  • Page 16 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Worksheet (Report) Pages ....... . 4 - 25 Probe/Connectors Usage .
  • Page 17 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview............5 - 1 Purpose of Chapter 5 .
  • Page 18 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL ECG Preamplifier ........5 - 18 MOD (Magneto-Optical Drive) .
  • Page 19 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Mechanical Concept and Overview ......5 - 42 Major Functions of CPN6 .
  • Page 20 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of Chapter 6 .
  • Page 21 No Connection to the Network at All ......7 - 22 GE remote service connection ......7 - 22 Error Messages .
  • Page 22 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
  • Page 23 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Tools ..........8 - 18 Backup all Exams of the “old”...
  • Page 24 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL CHAPTER 9 Renewal Parts Overview ............9 - 1 Purpose of Chapter 9 .
  • Page 25 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview............10 - 1 Periodic Maintenance Inspections .
  • Page 26 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Generic Procedure ........10 - 15 Data Sheet for Chassis Source Leakage Current .
  • Page 27 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Voluson® 730Pro / 730ProV scanner.
  • Page 28 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 1-1-3 Typical Users of the Basic Service Manual • Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation)
  • Page 29 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
  • Page 30 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 31 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment.
  • Page 32 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION "CAUTION - Dangerous voltage" (the lightning flash with arrowhead in equilateral triangle) is Rear side of Monitor used to indicate electric shock hazards.
  • Page 33 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION This symbol indicates that waste electrical and electronic equipment must not be disposed of...
  • Page 34 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 1-3-3 Mechanical Safety (cont’d) Keep the heat venting holes on the monitor unobstructed to avoid overheating of the monitor. CAUTION NOTE: Special care should be taken when transporting the unit in a vehicle: •...
  • Page 35 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 1-3-5 Labels Locations Figure 1-1 Labeling 1.) Caution Label 2.) CE-Label 3.) UL-Label 4.) Identification plate 5.) CW-Doppler (only if the CW-Doppler option is installed) 6.) Homologation label (for Japan only)
  • Page 36 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 1-3-7 Lockout/Tagout Requirements (For USA Only) Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the electrical Mains plug.
  • Page 37 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 1-4-2 Compliance The Voluson® 730Pro / 730ProV unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
  • Page 38 4855 W. Electric Avenue Milwaukee, WI 53219 1-262-524-5698 Customer Answer Center Fax: +1-414-647-4125 Europe Tel: +43 (0)7682 3800-26 GE Medical Systems Kretztechnik GmbH & Co OHG Fax: +43 (0)7682 3800-47 Service Department - Ultrasound Tiefenbach 15 A-4871 Zipf Austria Tel: +33-13083-1300...
  • Page 39 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 2 Pre-Installation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Voluson®...
  • Page 40 Electrical Requirements NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 41 2-2-2-5 Unit Power Plug If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. Chapter 2 - Pre-Installation...
  • Page 42 Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit RF RF gaskets touch metal leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, cable and peripherals length, material, and routing are all important;...
  • Page 43 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Operation: Ambient temperature 18° to 30° C Storage: -10° to 50° C NOTE: Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
  • Page 44 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 45 Emergency oxygen supply • Storage for linens and equipment • Nearby waiting room, lavatory, and dressing room • Dual level lighting (bright and dim) • Lockable cabinet ordered by GE for its software and proprietary manuals Chapter 2 - Pre-Installation...
  • Page 46 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 2-3-4 Minimal Floor Plan Suggestion Figure 2-1 Minimal Floor Plan Section 2-3 - Facility Needs...
  • Page 47 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of the DICOM Network Function DICOM (Digital Imaging and Communications in Medicine) services provide the operator with clinically useful features for moving images and patient information over a hospital network.
  • Page 48 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 2-10 Section 2-3 - Facility Needs...
  • Page 49 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are procedures to receive, unpack and configure the equipment.
  • Page 50 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-2-2 Installation Warnings 1.) Since the Voluson® 730Pro / 730ProV weighs approximately 136 kg (300 lb.) without options, two people are required to unpack it. Two people are required for installing any additional items in excess of 16kg / 35 pounds (e.g., Monitor).
  • Page 51 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE UNIT! Two people should unpack the unit because of its weight.
  • Page 52 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment CAUTION Transport only with forklift or stracker truck. During transport pay attention to the point of gravity (“tilt and drop” indicator)! Have two people available to unpack the Voluson®...
  • Page 53 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 3-4 Unpacking Procedure Step Task Open front panel by removing the screws from the side panels. Remove probes and accessories from the wooden box.
  • Page 54 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order.
  • Page 55 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-5 Connection of Auxiliary Devices NOTE: Normally the auxiliary devices and peripherals come already installed with the system. GES -...
  • Page 56 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-5 Connection of Auxiliary Devices (cont’d) 1.) Carefully remove plastic caps using a knife and loosen screws and washers (see: Figure 3-6).
  • Page 57 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-1 Monitor Connection NOTE: The monitor comes already installed with the system. 3-5-2 Foot Switch Connection NOTE: To adjust the Foot Switch, refer to the Voluson® 730Pro / 730ProV Basic User Manual, Chapter 17.3.3, System Setup - Peripherals.
  • Page 58 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-3 ECG-preamplifier Connection Figure 3-9 ECG Connection Scheme 3-10 Section 3-5 - Connection of Auxiliary Devices...
  • Page 59 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-4 Global Modem Connection Figure 3-10 Global Modem Connection Scheme Never install the telephone wiring, or use the modem during an electric storm;...
  • Page 60 3-5-5 S-VHS Video Recorder Connection 3-5-5-1 Mitsubishi HS-MD3000 ONLY the specific GE - Kretztechnik Version of this video recorder type may be used in CAUTION connection with the system! Figure 3-11 Mitsubishi - Video Recorder Connection Scheme Please use the proper connection set and remote cable.
  • Page 61 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-5-2 Sony SVO-9500MD Figure 3-12 Video Recorder Connection Scheme Set DIP Switches on back of the VCR as shown: Switch 1:...
  • Page 62 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-6 B/W Video Printer Connection Figure 3-14 B/W Video Printer Connection Scheme Set DIP Switches on back of the Black/White printer as shown in the image (all DOWN).
  • Page 63 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-7 Line Printer Connection Figure 3-15 Line Printer Connection Scheme Please observe that the Line Printer (HP 990cxi or HP 995c) has to be located outside of the CAUTION patient environment (acc.
  • Page 64 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-8 Digital Color Printer Connection Figure 3-16 Digital Color Printer Connection Scheme Pay attention to lateral distances. See Instruction Manual of the printer! CAUTION The Printer Supply Voltage must be the same as the Output Voltage of the Voluson®...
  • Page 65 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-9 Bluetooth Printer Connection Figure 3-17 Bluetooth Printer Connection Scheme Please observe that the complete Bluetooth Printer Assembly has to be located outside of the CAUTION patient environment (acc.
  • Page 66 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-10 External USB-Devices 3-5-10-1 Connection When an external USB-storage device (such as a USB-memory stick or an external hard disk) is connected to the Voluson®...
  • Page 67 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-5-10-2 Disconnection (cont’d) To stop the external device, select it and click the STOP button. A dialog shows which components will be stopped. To finish the process, click [OK].
  • Page 68 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-6 Completing the Installation 3-6-1 Power On / Boot Up 3-6-1-1 Scanner Power On 1.) Connect the Power Cable to the back of the system.
  • Page 69 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-6-2 Power Off/ Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 70 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-6-3 Transducer Connection NOTE: Prior to connecting or disconnecting a probe, freeze the image. It is not necessary to turn OFF power to connect or disconnect a transducer.
  • Page 71 The Bluetooth Printer Connection Set as well as the color printer CAUTION (HP 5600 Series or Olivetti Job Jet 210) MUST NOT be installed by the user! For installation please contact your local distributor or GE service representative. Chapter 3 - Installation 3-23...
  • Page 72 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-1 Installing Line Printer HP 990cxi or HP 995c 1.) Power off/Shutdown the system as described in: Section 3-6-2 on page 3-21.
  • Page 73 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-1-1 Install the HP 990cxi or HP 995c printer software/driver (cont’d) 5.) Click the NEXT buttons to start the Hardware Wizard and to locate the driver files.
  • Page 74 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-2 Installing Digital Color Printer Sony UP-D21MD or UP-D23MD 1.) Power off/Shutdown the system as described in: Section 3-6-2 on page 3-21.
  • Page 75 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-2-1 Install the UP-D21MD / UP-D23MD printer software/driver (cont’d) 6.) Use the BROWSE button to search the following path (see: Figure 3-27) •...
  • Page 76 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-3 Printer Installation manually 1.) On the control panel, press the UTILITIES key. 2.) In the “Utilities” menu on the left side of the screen, select SYSTEM to invoke the setup desktop.
  • Page 77 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-3 Printer Installation manually (cont’d) 6.) Click the ADD PRINTER button. A warning message appears: Please read this message carefully and click YES if you have skills to do this.
  • Page 78 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-3 Printer Installation manually (cont’d) 8.) Select the corresponding Printer Port (e.g., Figure 3-36 = USB001) and click NEXT. Figure 3-36 Select Printer Port 9.) In the following window select the HAVE DISK button.
  • Page 79 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-3 Printer Installation manually (cont’d) Figure 3-38 Select Printer Driver path (C:\Utilities\PrinterDriver\..) 11.)Click OPEN. 12.)Verify the selected Printer Driver path and confirm with OK.
  • Page 80 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-3 Printer Installation manually (cont’d) 13.)Select the manufacturer and model of your printer and confirm with the NEXT button. 14.)Assign a name, decide if the printer should be used as default printer and confirm with NEXT.
  • Page 81 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-4 Adjustment of Printer Settings 1.) After system restart, press the UTILITIES key, and then select SYSTEM from the menu area.
  • Page 82 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-4-1 HP 990cxi / HP 995c- Printer Settings 1.) Call up the ‘Printer Properties’; operation see: Section 3-7-4 "Adjustment of Printer Settings"...
  • Page 83 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-4-2 UP-D21MD / UP-D23MD - Printer Settings 1.) Call up the ‘Printer Properties’; operation see: Section 3-7-4 "Adjustment of Printer Settings"...
  • Page 84 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-4-2 UP-D21MD / UP-D23MD - Printer Settings (cont’d) 3.) Select the page. From the “Color Adjust” pop-up menu select: GRAPHICS a.) Color Balance: Cyan = 0; Magenta = 0; Yellow = 0 b.) Gamma Select: Gamma 1...
  • Page 85 The Bluetooth Connection Set as well as the color printer MUST NOT be installed by the user! CAUTION For installation please contact your local distributor or GE service representative. 1.) Call up the ‘Printer Properties’; operation see: Section 3-7-4 "Adjustment of Printer Settings"...
  • Page 86 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-7-5 Printer Remote Control Selection To assign the Remote PRINT A key, Remote PRINT B key and Report Printer to desired Printer: 1.) On the control panel, press the UTILITIES key.
  • Page 87 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-8 System Configuration 3-8-1 System Specifications 3-8-1-1 Physical Dimensions of Voluson® 730Pro / 730ProV The physical dimensions of the Voluson® 730Pro / 730ProV unit are summarized in Table 3-5.
  • Page 88 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-8-3 System Setup Modifications of system parameters are supported by diverse dialog pages and windows on the system setup desktop: •...
  • Page 89 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-8-3-1 How to enter Date and Time Select the page in the System Setup see: Figure 3-49. GENERAL 1.) Select the “Date Format” (only one can be active).
  • Page 90 How to change Video Norm BT Version: at BT´04 systems: Changing the Video Norm is not possible by the user. Please contact your GE service representative. BT Version: at systems that were upgraded to BT´05: 1.) Press the UTILITIES key on the Control Panel and then select the SYSTEM item from the menu.
  • Page 91 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-8-4 On-Board Optional Peripherals Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories. All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer.
  • Page 92 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-8-5 External I/O Connection Panel (GES) Figure 3-50 External I/O Panel Connectors Table 3-9 External I/O Connector Descriptions Item Connector Name...
  • Page 93 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-8-5-1 External I/O Pin Outs Table 3-10 VGA OUT Connector, Sub-D 15 Pin Pin No Output Signal Description VGA OUT1 R...
  • Page 94 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 3-14 COMP VIDEO OUT Connector Pin No Signal Description 1 (Center Pin) Signal 2 (Coax Pin) Signal GND Table 3-15...
  • Page 95 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Electrical Specifications on VGA Connector • Signal Level: 700 mV at 75 Ohm • H/V Sync: TTL Level Table 3-17 Video Specifications for Composite, Video Connector...
  • Page 96 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-10 Software/Option Configuration Refer to the Voluson® 730Pro / 730ProV Basic User Manual, Chapter 17.3.1, System Setup - General, for information on configuring items like Clinic Name, Language, Display, Date/Time, Date Format and Time Format.
  • Page 97 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-11 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: Type: Serial #: REV: CONTACT INFORMATION Name Title...
  • Page 98 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 3-12 Network IP Address Configuration NOTE: Following Information must be provided by customer or hospital engineer before you can start: A Station name, AE Title, IP address and Port Number for the Voluson®...
  • Page 99 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 3-12-1 Map Network Drive 1.) Select the MAP NETWORK DRIVE button (in the System Setup - Network page, see: Figure 3-51) to open a dialog where the system can be connected to a shared network drive of another server.
  • Page 100 Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. 3-52 Section 3-13 - Paperwork...
  • Page 101 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the Voluson® 730Pro / 730ProV scanner diagnostics by using the built-in service software, and power supply adjustments.
  • Page 102 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place.
  • Page 103 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-3-1-1 Scanner Power On (cont’d) 5.) Press the ON/OFF Standby switch left below the control panel. ON/OFF Standby switch Figure 4-2 ON/OFF Standby Switch When the ON/OFF Stand-By switch left below the control panel is pressed, the System (including the Back-end Processor) starts and the software code is distributed to initiate the scanner.
  • Page 104 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-3-3 System Features 4-3-3-1 Control Panel Figure 4-4 Control Panel Tour 1.) Ultrasound scanning gel holder 2.) Sonoview (Image Management) key 3.) VCR Remote Control key...
  • Page 105 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-3-3-2 Menu Control Figure 4-5 Menu Control Area- 2D Main Menu 1.) User Program: shows all setting for the active application. The active one is highlighted 2.) Additional functions which are supported by the selected Mode.
  • Page 106 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-3-3-3 Monitor Display Figure 4-6 Monitor Display Tour Table 4-2 Monitor Display Features 1. Logo 14. Image Area 2. Patient Name (Last-, First-, Middle Name) 15.
  • Page 107 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 4-4 Functional Checks For a basic functional check of the system’s different modes, following pages will familiarize you with image optimization for: •...
  • Page 108 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-1 2D Mode Checks Figure 4-7 2D Main and 2D Sub Table 4-3 2D Mode Functions Step Task Expected Results Rotate the 2D MODE key to adjust the sensitivity (brightness) of the entire 2D Mode Gain image.
  • Page 109 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 4-3 2D Mode Functions Step Task Expected Results Speckle Reduction Imaging is a smoothing type filter to reduce speckle in the (Speckle Reduction Imaging) ultrasound image.
  • Page 110 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 4-3 2D Mode Functions Step Task Expected Results Enhance brings out subtle tissue differences and boundaries by enhancing ENHANCE the gray scale differences corresponding to the edges of structures.
  • Page 111 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-2 M Mode Checks Figure 4-8 M Main and M Sub Table 4-4 M Mode Functions Step Task Expected Results Cursor Position Adjust the M Cursor position with the TRACKBALL in the 2D Single image.
  • Page 112 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-2-1 MCFM Mode Check By pressing the M control and the C control, the MCFM mode is switched on in the preparation mode.
  • Page 113 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-3 Spectral Doppler Mode Checks NOTE: Different menus are displayed depending on which Spectral Doppler Mode (PW or CW) is selected.
  • Page 114 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 4-5 Spectral Doppler Mode (PW, CW) Functions Step Task Expected Results The Velocity Range display is governed by the pulse repetition frequency (PRF) Exceeding the maximum PRF, the HPRF-Mode is automatically switched on.
  • Page 115 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-4 Color Doppler Mode Checks NOTE: Different menus are displayed depending on which Color Doppler Mode (CFM, PD or TD) is selected.
  • Page 116 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 4-6 Color Doppler Mode (CFM, PD, TD) Functions Step Task Expected Results DISPLAY MODE To select the CFM- Display Mode (V; V-T; V-Pow; Pow-T; or T).
  • Page 117 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-5 Volume Mode Checks Real Time 4D (freeware on Voluson® 730ProV), RT_4D_Biopsy*, VCI*, STIC*, TUI* and VOCAL II* are NOTICE Options. If these options is not a part of the system configuration, the appendant checks can be omitted.
  • Page 118 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 4-7 Pre-Volume Mode Functions Step Task Expected Results 4D REAL TIME Real Time 4D - continuous volume acquisition and parallel calculation of 3D rendered images...
  • Page 119 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-5-2 Functions after the 3D Acquisition (cont’d) Table 4-8 Functions after the 3D Acquisition Step Task Expected Results X-ROT : Rotation about X-axis of the reference image.
  • Page 120 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-5-3 Sub Menu Figure 4-14 3D/4D Sub Menu Table 4-9 Sub Menus Step Task Expected Results To select the desired Render View Direction (green line symbolizes direction of the view).
  • Page 121 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-6 Using Cine 4-4-6-1 Activating Cine Press FREEZE, then roll the TRACKBALL to display the images of the stored sequence one by one.
  • Page 122 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-7 Generic Measurements NOTE: Different menus are displayed depending on: • the currently selected Application and Display Mode, • the selected “Study”, •...
  • Page 123 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-7 Generic Measurements (cont’d) 4-4-7-2 Circumference/Area Measurements 1.) Press the CALIPER key and then select the GENERIC AREA in the menu area.
  • Page 124 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-7 Generic Measurements (cont’d) 4-4-7-4 Measurements in Spectral Doppler Mode NOTE: The Spectral Doppler image is displayed based on time (X-axis) and velocity (Y-axis).
  • Page 125 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-8 Calculations The Voluson® 730Pro / 730ProV system supports calculation packages and application-oriented patient Worksheets (Reports) for following applications: • Abdomen Calculations •...
  • Page 126 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-9 Probe/Connectors Usage 4-4-9-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case.
  • Page 127 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-10 Image Management (Sonoview) For Sonoview - Image Management functionality refer to Chapter 15 in the Basic User Manual of Voluson® 730Pro / 730ProV. It talks about several topics: •...
  • Page 128 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-11 Using the MOD (Magneto-Optical Drive) The MOD (Magneto-Optical Drive) is an Option. NOTICE The optional 3.5 inch Magneto-Optical disk drive supports the following medias: 1.3GB;...
  • Page 129 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-11-1 Formatting Media To erase/format the backup media, DVD/CD+(R)W or MOD, press the SONOVIEW key on the Control panel. The Sonoview screen appears on the monitor; see: Figure 4-15 on page 4-27.
  • Page 130 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-4-11-1 Formatting Media (cont’d) By selecting DVD/CD the unit displays the “Erase DVD/CD” window as shown in Figure 4-19. 1.) Select the “Erase Mode” and click the [OK] button to start the process.
  • Page 131 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 4-5 Backup and Restore Database, Preset Configurations and Images 1.) Press the UTILITIES key on the control panel. The menu area changes to the Utilities menu.
  • Page 132 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-1 Save User Settings Only (Application Settings) The User Settings contains: • User Programs • Auto Text • 3D/4D Programs 1.) Insert a DVD/CD+(R)W or MO (Magneto-Optical Disk) into the drive.
  • Page 133 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-2 Load User Settings Only (Application Settings) The loading procedure overwrites the existing application settings on the local hard drive. Make sure to CAUTION insert the correct DVD/CD or MO.
  • Page 134 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-2 Load User Settings Only (Application Settings) (cont’d) 7.) Select the Complete Backup (marked blue; see: Figure 4-25) and click the >> button to copy the Complete Backup into the Load Data field.
  • Page 135 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-3 Save Full Backup (Presets, Configurations & Application Settings) A full backup always contains the following data • Patient demographic and exam data (database containing the patient data and measurements) •...
  • Page 136 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-3 Save Full Backup (Presets, Configurations & Application Settings) (cont’d) When the Full Backup is stored on a network drive (to map a network drive see: Section 3-12-1 "Map...
  • Page 137 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) Always backup any data before an upgrade; see: Section 4-5-3 on page 4-35.
  • Page 138 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) (cont’d) 1.) Press the UTILITIES key on the control panel. The menu area changes to the Utilities menu.
  • Page 139 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) (cont’d) 8.) Click the NEXT button and then select YES to start, or NO to cancel the restore procedure.
  • Page 140 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-5-6 Archiving Images 1.) Press the SONOVIEW key on the control panel. 2.) Insert the DVD/CD+(R)W or MOD into drive. If required, format/erase the media, see Section 4-4-11-1 "Formatting Media"...
  • Page 141 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 4-6 Software Configuration Checks Press the UTILITIES key and select the SYSTEM item from the “Utilities” menu - which is shown in the menu area on the left side of the screen.
  • Page 142 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 4-7 Peripheral Checks Check that peripherals work as described below: Table 4-12 Peripheral Checks Step Task to do Expected Result(s) Press the FREEZE key.
  • Page 143 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 4-8 Mechanical Function Checks 4-8-1 Rotation of the Control Console locking lever for locking and unlocking the control console Figure 4-37 locking lever under Control Console...
  • Page 144 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 4-9 Site Log 4-9-1 Site Log - System (Service Database) 1.) Press the UTILITIES key on the Control Panel. The menu area changes to the Utilities menu.
  • Page 145 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4-9-2 Site Log - Paper Documentation Table 4-16 Site Log Date Service person Problem Comments Chapter 4 - Functional Checks 4-45...
  • Page 146 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 4-46 Section 4-9 - Site Log...
  • Page 147 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains Voluson® 730Pro / 730ProV’s system concepts, component arrangement, and subsystem function.
  • Page 148 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-2 General Information Voluson® 730Pro / 730ProV is a digital beamforming curved-, linear- and phased array ultrasound imaging system. It has provisions for analog input sources like ECG and Phono.
  • Page 149 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-2 General Information (cont’d) Monitor without GEU Figure 5-2 Major System Components Major System Components: • GEF: Main Board Chassis:...
  • Page 150 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-2 General Information (cont’d) Monitor User-Interface-Module Trolley (housing, mechanical framework). GEM- External IO ECG+Diskdrive The following components Module are part of GW:...
  • Page 151 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-2 General Information (cont’d) The Voluson® 730Pro / 730ProV used digital beamforming technology which provides high resolution and high penetration performance. It is a general purpose, mobile, software controlled diagnostic ultrasound scanner.
  • Page 152 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-2 General Information (cont’d) Signal flow travels from the Probe Connector Panel to the Front End Electronics, to the Back-End Processor, and finally displayed on the monitor and peripherals.
  • Page 153 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-1-3 Color Doppler Mode Color Doppler is used to detect motion presented as a two-dimensional display. There are three applications of this technique: •...
  • Page 154 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-2 3D Imaging The Voluson® 730Pro / 730ProV Ultrasound System will be used to acquire multiple, sequential 2D images which can be combined to reconstruct a three dimensional image. These 3D images are useful in visualizing three-dimensional structures, and in understanding the spatial or temporal relationships between the images in the 2D sequence.
  • Page 155 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-3 Block diagram Voluson® 730Pro / 730ProV Figure 5-5 Top Level Architecture Chapter 5 - Components and Functions (Theory)
  • Page 156 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-3 Block diagram Voluson® 730Pro / 730ProV (cont’d) Figure 5-6 Voluson® 730Pro / 730ProV - Block diagram 5-10 Section 5-2 - General Information...
  • Page 157 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-4 Dataflow Control Description This section describes functions of Voluson® 730Pro / 730ProV Boards vs different Operation Modes. • CPR - Beam former Board •...
  • Page 158 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-4-1-1 Special B Mode Techniques a.) HI (Coded Harmonic Imaging) In one method of HI the RX-Frequency is doubled, so that the radial resolution is increased due to the higher RX-Frequency.
  • Page 159 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-4-3 D-Mode (Pulsed Wave- and Continuous Wave Doppler) 1.) CRS PRF-generator; see: 5-2-4-1 B-Mode After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF.
  • Page 160 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-4-5 CFM-Mode (Color Flow Mode) 1.) CRS PRF-generator; see: 5-2-4-1 B-Mode After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF.
  • Page 161 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-4-10 Sonoview write mode (store Image to Sonoview) 1.) CRS Not used 2.) CPR Not used 3.) CKV - DMA section Not used 4.) CKV - Video section...
  • Page 162 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-5-2 DICOM Voluson® 730Pro / 730ProV software package providing following DICOM functionality: • Storage Service Class • Print Management Service Class •...
  • Page 163 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-5-7 STIC (Spatio-Temporal Image Correlation)* With this acquisition method the fetal heart or an artery can be visualized in 4D. It is not a Real Time 4D technique, but a post processed 3D acquisition.
  • Page 164 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-6 Description of Hardware Options Table 5-3 Hardware Options HW-Options Description CW-Doppler 5-2-6-1 CW - Continuous Wave Doppler ECG Digital Module...
  • Page 165 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-2-6-5-1 Location in the Unit Figure 5-8 Modem (placed on the GEF-box) 5-2-6-5-2 LEDs The Modem has 10 LEDs on its front.
  • Page 166 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-3 Main board Chassis GEF Module The GEF Module contains the Front End processor and the Back End processor and the Secondary Power supply for the full GEF Chassis.
  • Page 167 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-4 FrontEnd Processor Figure 5-11 FrontEnd Processor - Block diagram Chapter 5 - Components and Functions (Theory) 5-21...
  • Page 168 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-4-1 FrontEnd - Board Descriptions 5-4-1-1 CPU - Probe Connector Board • 1 CW-Probe Connector • 3 Probe-Connectors 260pin • 1 Dummy-Probe Connector 260pin •...
  • Page 169 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-4-1-3 CPD-Beamformer-Sub-board 1x : Beamformer-ASIC generates TX-Pulses + TX-Focus +TX-Apodization, Rx Focus Delay and Summation 16x : TX-Power Amplifier 4x : TGC + Anti Aliasing Filter + Analogue Digital Conversion (ADC)
  • Page 170 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-4-1-4 CRW - CW-Doppler Board (optional) • Separate Clock Oscillator + Clock generator PRF, FN • TxDelayMatrix • Transmitter, • Receiver Amplifiers (RxAmp), •...
  • Page 171 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-4-1-7 CRS - Signal Processor Board A.) MID-Processor: 1.) For B, C, D modes: Multi Beam-DeInterleave Subtraction Filter (for HI) Digital TGC, DC-Cancelling...
  • Page 172 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL + FAN (10 -24V / 15 Watt; adjustable by software) • +TX_POW (+/-90V) • -TX_POW (+/-90V) • +TX_POW2 • -TX_POW2 Figure 5-14 Block diagram CPP...
  • Page 173 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-5 BackEnd Processor Figure 5-15 BackEnd Processor - Block diagram Chapter 5 - Components and Functions (Theory) 5-27...
  • Page 174 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-5-1 Block diagram CKV DMA Controller PCI Bus 32 Bit / 33MHz PLX PCI 9054 DMA 0 DMA 1 Host PC...
  • Page 175 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-5-2 BackEnd - Board Descriptions 5-5-2-1 SBC - Single Board Computer Built in or external Components: AGP VGA 100 Mbit LAN USB 2.0...
  • Page 176 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-5-2-7 CPP- Power Supply Secondary Board + Motor Power stage used for Supply of both FrontEnd and BackEnd DC-DC-Converter: 59VDC to following output voltages: + 3.3V, +/-5V, +12V, +/-15V,...
  • Page 177 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-6 Internal I/O StdbySwitch (User PowerON/OFF) GES - External IO Board GEM - Disk Drive Module SUPPLY DVD + (R)W...
  • Page 178 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-6 Internal I/O (cont’d) Figure 5-18 Connectors on the Main Module of the GEF Item Connector Name Description VGA1 Connector for the internal color video monitor...
  • Page 179 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-6 Internal I/O (cont’d) Figure 5-19 Connectors on PC-part of the GEF Item Connector Name Description USB-GEM Connector for Disk Drive Module GEM...
  • Page 180 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-7 Top Console The Voluson® 730Pro / 730ProV Operator Control Panel (OCP) consists of the following electronic subassemblies and/or functional components: •...
  • Page 181 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-7 Top Console (cont’d) Figure 5-21 Voluson® 730Pro / 730ProV - Control Console Chapter 5 - Components and Functions (Theory)
  • Page 182 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 5-5 Voluson® 730Pro / 730ProV - key codes key code Description Voluson® 730Pro / 730ProV - Functionality call-up of the patient data entry menu...
  • Page 183 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 5-5 Voluson® 730Pro / 730ProV - key codes key code Description Voluson® 730Pro / 730ProV - Functionality DOWN - penetration depth of the B-image...
  • Page 184 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 5-5 Voluson® 730Pro / 730ProV - key codes key code Description Voluson® 730Pro / 730ProV - Functionality UP - Soft Flip 2 (Toggle switch function)
  • Page 185 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-8 Monitor RIGHT LEFT KEY FUNCTION a. MENU KEY This button will enable the On Screen Display. This button is also used to select the function in the Main Menu or to save the setting in the Sub Menu.
  • Page 186 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-9 External I/O Figure 5-23 External I/O Panel Connectors Table 5-6 External I/O Connector Description Item Connector Name Description VGA OUT...
  • Page 187 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-10 Peripherals 5-10-1 General Information The GEM Module contains the DVD/CD+(R)W Drive. Additionally the Magneto Optical Drive and the ECG-preamplifier (MAN6) can be installed as an Option.
  • Page 188 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-11 Power Distribution Power Distribution depends on the currently installed CPN - Primary Power Module. NOTICE • Section 5-11-1 "CPN6 - Primary Power Module" on page 5-42 •...
  • Page 189 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-11-1-3 Fuses on Rear Panel (CPN6) F1: The main Input voltage is fused by a magnetic Circuit breaker (Rated current 16A) built in the Main Power switch labelled F1 F2:The AC Output voltage (115/230V) is fused by F2 (magnetic Circuit breaker 2.5/5A)
  • Page 190 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-11-2 CPN80 - Primary Power Module Figure 5-27 Primary Power Module - CPN80 5-11-2-1 Mechanical Concept and Overview The AC Power’s main tasks are to supply the various internal subsystems with AC power and to galvanically isolate the scanner from the on site Mains Power System.
  • Page 191 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-11-2-3 Fuses on Rear Panel (CPN80) F1 + F2: The main Input voltage is fused by two 16 Ampere fuses (rated current 16A) labeled F1 and F2.
  • Page 192 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-12 Mechanical Descriptions 5-12-1 Physical Dimensions 607,5 399,5 560,5 Figure 5-30 Physical Dimensions 5-46 Section 5-12 - Mechanical Descriptions...
  • Page 193 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-12-2 Monitor • Tilt: 11° forwards and backwards • Swivel: +/-90° rotation. 5-12-3 Top Console Positioning 30° Figure 5-31 Top Console Positioning...
  • Page 194 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-12-5 Assembly Drawing GW & GEU & Monitor Figure 5-33 Assembly Drawing GW & GEU & Monitor 5-48 Section 5-12 - Mechanical Descriptions...
  • Page 195 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-13 Air Flow Control 5-13-1 Air Flow Distribution The fans at the right side of the Main Board Chassis draw air into the scanner, through the filter grid, and pushes it through the card rack.
  • Page 196 Introduction The Service Platform will increase service productivity and reduce training and service costs. This web-enabled technology provides linkage to e-Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Service Platform contains: • specific software/hardware test modules, system setup, update, etc. for Voluson systems; see: Section 5-15 "Service Page"...
  • Page 197 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-14-3 iLinq Interactive Platform Many of the services of the Service Platform come from its integration with iLinq. If an iLinq checkout was performed on the Voluson® 730Pro / 730ProV, the ILINQ selection item is available in the menu area (on the left side of the screen) of the “Utilities”...
  • Page 198 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 5-15 Service Page 5-15-1 Introduction The Service Page contains specific software/hardware test modules, system setup, update, etc. for Voluson systems only.
  • Page 199 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 4.) Enter the password and click the ACCEPT button to display the Service Tools window. Auto Tester Monitor Test Printer Service Viewer...
  • Page 200 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 5-15-3-3 Service Viewer Provides common information about System Temperature, Probes, Working hours of system components and probes. 1.) Select the SERVICE VIEWER button to get access to the E-Service page.
  • Page 201 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 1.) Click the MAINTENANCE REPORT button. The following message box will be displayed. Figure 5-41 Maintenance Report 2.) Fill in the requested information and click OK.
  • Page 202 Installation of the Line Printer is possible without entering the Windows Desktop. Operation see: 3-7-3 "Printer Installation manually" on page 3-28. Only accessories explicitly recognized and released by the system manufacturer GE WARNING WARNING Medical Systems - Kretztechnik may be used in connection with the system.
  • Page 203 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
  • Page 204 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 6-3 Monitor Adjustment The Monitor can be tilted or rotated. • Tilt angle: up/down 11° • Rotation angle: right/left 90°...
  • Page 205 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 6-3-1-1 Degauss Degaussing refers to the process of removing magnetic-field effects from the monitor. Operation of the monitor within a magnetic field may adversely effect color purity. Degaussing can be used to correct the problem.
  • Page 206 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 6-3-1-3 Contrast and Brightness Adjusting the monitor´s contrast and brightness is one of the most important factors for proper image quality. If these controls are set incorrectly, the Gain, TGC, Dynamic Range and even Acoustic Output may have to be changed more often than necessary to compensate.
  • Page 207 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 6-4 Control Console, Transport Lock 6-4-1 Control Console The control console can be rotated 30° to the right. When rotating the control console grasp it only the front grip.
  • Page 208 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 6-5 Trackball Adjustment Adjustment of the mechanical movement may be necessary to ensure smooth running of the trackball. 1.) Remove the outer fixing ring by turning it counterclockwise.
  • Page 209 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 210 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-2 Collect Vital System Information The following information is necessary in order to properly analyze data or images being reported as a...
  • Page 211 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-2-1 Needed Data - in case of System crash or Errors In case of a system crash, or errors that are probably software bugs, provide as much detail information NOTICE as possible before contacting the OLC (On-Line Center).
  • Page 212 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-3 Check Points Voltages 7-3-1 How to check power Open the locking, flap the service lid on the back of theVoluson® 730Pro / 730ProV upwards, and check the status of the Diagnostic LED´s.
  • Page 213 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-4 Screen Captures and Logs There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center.
  • Page 214 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-4-2-1 Export System Data (by pressing the ALT + D key) ALT + D uses “Full Backup” to gather data from the system. In addition it creates one screenshot (Alt-D.bmp) of the point in time when ALT + D was pressed.
  • Page 215 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-4-2-2 Export Log´s and System Data (via Service Page) 1.) Press the UTILITIES key, select the SYSTEM item from the menu area.
  • Page 216 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-5 How to use the Auto Tester program 1.) Press the UTILITIES key and select the SYSTEM from the menu area on the left side of the screen.
  • Page 217 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-5 How to use the Auto Tester program (cont’d) The following message box apppears. Pause key on the right upper corner...
  • Page 218 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-5 How to use the Auto Tester program (cont’d) The following screen will appear. “DVD/CD Burn” icon Figure 7-9 Auto Tester Finished 10.)Insert an empty DVD/CD+(R)W in the Drive and select the “DVD/CD Burn”...
  • Page 219 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-6 Minimum Configuration to Boot/Scan 7-6-1 Booting without using the Ultrasound Hardware If the low-level ultrasound hardware is failing, the application can not be initialized correctly.
  • Page 220 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-6-2 Minimum Configuration to Scan Following cables must be connected to scan; see: Figure 7-14 a.) PS2 (connector for Mouse/Keyboard) * PS2 Adapter is ONLY used, when the currently installed PC-Board is an IBASE Standard Board Computer.
  • Page 221 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-7 Troubleshooting Trees and Instructions 7-7-1 System Does Not Power On / Boot Up Is the scanner plugged Make sure the...
  • Page 222 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-7-2 Noise disturbs the Image Noise is shown in Image. Is there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices.
  • Page 223 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-7-3 Trackball The trackball is low sensitive! Does the trackball move at all? Remove the ball from the Trackball unit, and clean it.
  • Page 224 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-7-4 System Does Not Power Off / Shutdown System has been switched off system switched off Primary Power (CPY System starts with the automatically in approx.
  • Page 225 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-7-5 Monitor Troubleshooting Fault symptom Check these items Check the power cord is properly connected. CPN6 only : Check the Power Switch of Peripherals ( F2 ) on back of system is set to the “ON”...
  • Page 226 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-7-6 Unable to Record to VCR Unable to record to VCR! Is the VCR properly connected? Check that all signal, and power supply cable connections to the VCR unit are OK.
  • Page 227 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-7-7 Printer Troubleshooting Unable to use Printer(s)! Is the printer(s) properly connected? Check that all signal, and power supply cable connections between Printer and Scanner are OK.
  • Page 228 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-7-7 Printer Troubleshooting (cont’d) Still unable to use the Printer(s)! Is printing unavailable even by manual operation? Check the signal cables between GEF and Backpanel.
  • Page 229 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7-7-8 DVD/CD+(R)W Troubleshooting (DVD/CD Drive) 1.) Insert an empty DVD/CD+(R)W into the Drive. 2.) Enter “Sonoview” by pressing the SONOVIEW key on the control panel; see Figure 7-22.
  • Page 230 3.) Connect a network cable between your Scanner and your PC. Try to ping from the Scanner to the IP address on the PC. If OK, the hardware connection inside the Scanner is OK. 7-7-10-2 GE remote service connection The customer gets visual information about the network status. The network status is defined as follows: BLACK - Cable disconnected or no network signal on a connected cable.
  • Page 231 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 7-8 Error Messages If the problem (error message) still exists after performing the described actions, call technical support. NOTICE Error Messages...
  • Page 232 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions B-DynContrast Out Of Range press ok and save this user-setting once again BF: can't set BM RxApod restart the system BF: can´t set C Rx Apod...
  • Page 233 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions CtrlMvGrpRepresentations::isInWriteMode not supported for restart the system mode nr xx CW quant TX Frequ out of range restart the system...
  • Page 234 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions Enhance Out Of Range press ok and save this user-setting once again Ensemble Out Of Range press ok and save this user-setting once again Error in CreateCineImage try to store again;...
  • Page 235 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions File Pos load volume files from other storage medium File Read Error load volume files from other storage medium...
  • Page 236 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions M_Gain Out Of Range press ok and save this user-setting once again M_Reject Out Of Range press ok and save this user-setting once again...
  • Page 237 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions Overflow restart the system Persistance Out Of Range press ok and save this user-setting once again Persistence coeff page index too big...
  • Page 238 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions RepresentationManager::pActualUnitWrtAddr called within restart the system Transaction!! RepresentationManager::pNextUnitsCompleted called within restart the system Transaction!! RepresentationManager::removeAllListener called within restart the system...
  • Page 239 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions RT_ConnectionMngr::vAssembleRTSet: no restart the system RT_TissueFilterBlock found for TISSUE3D Blocks RT_ECG_Block::bStart ECGInterface failed!! restart the system RT_MColorFilterBlock::RT_MColorFilterBlock: Memory...
  • Page 240 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Error Messages Actions Thickness mismatch xx - GIP xx restart 3D (go to 2D); restart the system UI_BBC_Wnd::vSet() has an wrong ImageType...
  • Page 241 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems.
  • Page 242 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-1-2 Returning/Shipping Probes and Repair Parts Equipment being returned must be clean and free of blood and other infectious substances. GEMS policy states that body fluids must be properly removed from any part or equipment prior to shipment.
  • Page 243 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-2-4 Preparations (cont’d) 1.) Press the UTILITIES key on the control panel once to display the Utilities menu. 2.) Select the SYSTEM item from the Utilities menu - which is displayed on the left side of the screen.
  • Page 244 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-2-4 Preparations (cont’d) 5.) Select the page on the System Setup desktop screen. NETWORK a.) Click on the DICOM / SONOVIEW CONFIGURATION button in the page.
  • Page 245 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-2-5 Software - Installation Procedure (via Service Page) 1.) Insert the System DVD into the drive. 2.) Press the UTILITIES key and then select SYSTEM item from the Utilities menu.
  • Page 246 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-2-5 Software - Installation Procedure (via Service Page) (cont’d) 6.) Confirm the following message with YES. Figure 8-7 Update UIS message 7.) A new window pops up on the screen and the InstallShield starts extracting files.
  • Page 247 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-2-5 Software - Installation Procedure (via Service Page) (cont’d) 9.) Complete the Software installation with the FINISH key. Figure 8-10 V730 Software Setup The system restarts automatically.
  • Page 248 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 8-3 User Settings Only (Application Settings) Loading Procedure 8-3-1 Introduction The User Settings contains: • User Programs • Auto Text •...
  • Page 249 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 8-5 Service Platform (SSW) Upgrade Procedure 8-5-1 Manpower One Person, 10 min. 8-5-2 Tools System DVD or Upgrade CD 8-5-3 Upgrade Procedure 1.) Insert the System DVD or Upgrade CD into the drive.
  • Page 250 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 8-5 Service Platform (SSW) Upgrade Procedure (cont’d) 8.) Select the DVD/CD DRIVE button 9.) Browse for the install.bat file and click OK.
  • Page 251 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 8-6 Electronic User Manual (EUM) Upgrade Procedure 8-6-1 Manpower One Person, 15 min. 8-6-2 Tools System DVD or Upgrade CD...
  • Page 252 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-6-4 EUM - Upgrade Procedure 1.) Click the EUM button. The “Update Software” dialog appears. 2.) Select the DVD/CD DRIVE button 3.) Browse for the EUMSetup_en_V730Pro.exe file and click OK.
  • Page 253 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-6-4 EUM - Upgrade Procedure (cont’d) 7.) Click the EUM button again to install the new EUM. 8.) Select the MO DRIVE or the DVD/CD DRIVE button, depending on the storage medium you use 9.) Browse for the EUMSetup_en_V730Pro.exe file again and click OK.
  • Page 254 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 8-7 Software and Functional Checks after the Upgrade 1.) Press the UTILITIES key on the control panel once to display the Utilities menu.
  • Page 255 Option page of Option page of System type: System type: Voluson 730Pro V Voluson 730Pro Figure 8-21 Option page in the System Setup • D = Demo (Option is activated for demo and expires on the date shown in the “Valid” column •...
  • Page 256 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-8-1-1 Operation for installing a “Demo Key” or a “Permanent Key”: 1.) Position the cursor inside the input field desired and press the key.
  • Page 257 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 8-9 Transfer of Patient Database and Images from System-to-System 8-9-1 Introduction It is possible to backup Sonoview Image data to a Mapped Network Drive (Z:\) NOTICE Please make sure that the server you are connecting to is trustworthy and reliable.
  • Page 258 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-9-2 Transfer of Patient Database and Images via Sonoview 8-9-2-1 Introduction This section describes how to transfer the complete patient database and images from one system (= “old”...
  • Page 259 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-9-2-4 Backup all Exams of the “old” system (cont’d) 7.) Click on the close button and then select the SETTINGS icon on the left side of the screen.
  • Page 260 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-9-2-5 Restore all Exams (of the “old” system) to the “new” system (cont’d) The folder which is selected to be renamed (e.g., A09008, that contains the backup data from a different...
  • Page 261 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 8-9-2-5 Restore all Exams (of the “old” system) to the “new” system (cont’d) 11.)Select the desired action: The exam on the hard disk is replaced with the exam in the backup.
  • Page 262 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 8-10 Replacement of the Monitor Task Lamp 8-10-1 Manpower One person, 15 min. 8-10-2 Tools Philips screwdriver 1 and 2, stubby (length ~30mm)
  • Page 263 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 8-11 Replacement of the Trackball top fixation ring 8-11-1 Manpower One person, 5 min. 8-11-2 Trackball top fixation ring - Replacement Procedure 1.) Remove the fixation ring by turning it counterclockwise.
  • Page 264 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 8-24 Section 8-12 - Replacement of Digipots and TGC Slider controls...
  • Page 265 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the Voluson® 730Pro / 730ProV.
  • Page 266 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-2 List of Abbreviations • CKV - Video Converter Board • CPD - Sub-Board on Beamformer (CPR) • CPE - Motherboard Extension (Backpanel I/O-Card) •...
  • Page 267 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-3 Parts List Groups Figure 9-1 Console Views Table 9-2 Mechanical and user accessible parts Item Part Group Name Table Number...
  • Page 268 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-4 Housing (GW) and additional Console Hardware SNO less/equal ≤ A32513 SNO greater/equal ≥ A32514 Figure 9-2 Housing (GW) Section 9-4 - Housing (GW) and additional Console Hardware...
  • Page 269 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-3 Housing (GW) Item Part Name Part Number Description Rear Handle for Trolley; Rear Handle for Trolley KTZ220031 used in serial production SNO less/equal ≤ A32513 (SNO ≤...
  • Page 270 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-4 Housing (GW) and additional Console Hardware (cont’d) Figure 9-3 Additional Console Hardware Table 9-4 Additional Console Hardware Item Part Name...
  • Page 271 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-5 User Interface (GEU Top Console) Figure 9-4 User Interface (GEU) Chapter 9 - Renewal Parts...
  • Page 272 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-5 User Interface (GEU) Item Part Name Part Number Description Voluson® 730Pro / 730ProV Front Handle KTZ220039 Front Handle User Interface (only used with GEU60A - KTZ154688) Voluson®...
  • Page 273 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-6 Monitor + Monitor replacement parts Figure 9-5 Monitor + Monitor replacement parts Table 9-6 Monitor + Monitor replacement parts...
  • Page 274 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-7 Disk Drives (GEM) 305 306 307 308 301 302 Figure 9-6 Disk Drives (GEM) Table 9-7 Disk Drives (GEM)
  • Page 275 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-8 Main Power Module (CPN) 414 415 410 411 412 Figure 9-7 Main Power Module (CPN) Chapter 9 - Renewal Parts...
  • Page 276 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-8 Main Power Module (CPN) Item Part Name Part Number Description CPN6 Power Supply Module KTZ195905 CPN6 Power Supply Module...
  • Page 277 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-9 Main Board Module (GEF) Figure 9-8 Main Board Module (GEF) Table 9-9 Main Board Module (GEF) Item Part Name...
  • Page 278 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 9-9-1 FrontEnd (US-Part) Figure 9-9 FrontEnd (US-Part) Table 9-10 FrontEnd (US-Part) Item Part Name Part Number Description KTZ195904 Beam former Board CPR82.P11 Beam former Board...
  • Page 279 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 9-9-1 FrontEnd (US-Part) (cont’d) Figure 9-10 FrontEnd (US-Part) cont´d Table 9-11 FrontEnd (US-Part) cont´d Item Part Name Part Number Description Electrical Signal- and Supply-Connection for all KTZ207456 CPK80c.P5 Motherboard V730...
  • Page 280 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 9-9-2 BackEnd Processor (PC-Part) Figure 9-11 Back Processor (PC-Part) 9-16 Section 9-9 - Main Board Module (GEF)
  • Page 281 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-12 Back Processor (PC-Part) Item Part Name Part Number Description KTZ207418 PC-Video converter Board CKV81.P6 Video management Board KTZ195902 Motherboard Extension CPE80.P5 Motherboard - Extension...
  • Page 282 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-10 Options and Upgrades 602 603 Figure 9-12 Options and Upgrades Table 9-13 Options and Upgrades Item Part Name Part Number...
  • Page 283 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-11 Miscellaneous Cables Table 9-14 Miscellaneous Cables Item Part Name Part Number Description Set contains two adapters: • Audio Adapter Line-IN/OUT...
  • Page 284 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description 12V/5V-Power Supply-Distributor-Cable (2x) for PC- Component (yellow-black, red-black) KTZ212426 Cable HDD Y-Power...
  • Page 285 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description External Rear Panel with electrical Signal- and Supply- Connection-Cables to the Voluson® 730Pro / 730ProV Main-Unit (internal) Rear-Panel.
  • Page 286 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description Power Cord Europe 230V/240V KTZ212317 Power Cord Europe 230V Power Cord Japan Hospital Grade KTZ212448 Power Cord Japan (Hosp.grade)
  • Page 287 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description Set contains two parts: USB for PC-Slot, Connector on Backpanel. Cables are connected to PC-Board.
  • Page 288 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description Set contains two cables: Cable TMDS (Digital Video Cable from SBC to CKV)
  • Page 289 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-12 Optional Peripherals and Accessories Figure 9-13 Optional Peripherals and Accessories Chapter 9 - Renewal Parts 9-25...
  • Page 290 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-15 Optional Peripherals and Accessories Item Part Name Part Number Description KTZ211332 NTSC/PAL Sony B/W Video Printer UP-895MD (CE) (H46801A)
  • Page 291 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-12 Optional Peripherals and Accessories (cont’d) Figure 9-14 Optional Peripherals and Accessories cont´d Table 9-16 Optional Peripherals and Accessories cont´d...
  • Page 292 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-17 System Manuals Item Part Name Part Number Kretz # Description Voluson® 730Pro / 730ProV Service KTZ105947 KTI105947 Manual System User Manuals Basic User Manual, Voluson®...
  • Page 293 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-13 Probes 9-13-1 2D-Probes Figure 9-15 2D curved array Transducers Table 9-18 2D curved array Transducers Item Part Name Part Number...
  • Page 294 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 9-13-2 9-13-12D-Probes (cont’d) Figure 9-16 2D linear- and phased array Transducers Table 9-19 2D linear- and phased array Transducers Item Part Name...
  • Page 295 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 9-13-3 Real-Time 4D Volume Probes Figure 9-17 Real-Time 4D Volume Probes Table 9-20 Real-Time 4D Volume Probes Item Part Name Part Number...
  • Page 296 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 9-13-3 Real-Time 4D Volume Probes (cont’d) Figure 9-18 Real-Time 4D Volume Probes Table 9-21 Real-Time 4D Volume Probes Item Part Name...
  • Page 297 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 9-13-4 CW-Pencil Probes Figure 9-19 CW- Pencil Probes Table 9-22 CW- Pencil Probes Item Part Name Part Number Description single element Continuous Wave (CW) Doppler pencil probe with a KTZ195540 PCW4.0...
  • Page 298 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 9-14 Biopsy Needle Guides Figure 9-20 Biopsy Needle Guides 9-34 Section 9-14 - Biopsy Needle Guides...
  • Page 299 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 9-23 Biopsy Needle Guides Item Part Name Part Number Description ***** disposable Biopsy needle guide for probe IC5-9H E8385MJ (disposable) needle diameter: <...
  • Page 300 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 9-36 Section 9-14 - Biopsy Needle Guides...
  • Page 301 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your Voluson® 730Pro / 730ProV system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
  • Page 302 Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
  • Page 303 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont’d) Table 10-2 Periodic Maintenance Schedule Item Service at Indicated Time...
  • Page 304 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Table 10-2 Periodic Maintenance Schedule Item Service at Indicated Time Notes Also after corrective • Console Leakage maintenance or as required by Current Checks your facilities QA program.
  • Page 305 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Periodic Maintenance...
  • Page 306 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary.
  • Page 307 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks (see also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary.
  • Page 308 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to Table 3-8, “Approved Peripherals,”...
  • Page 309 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-5-5 Physical Inspection Table 10-9 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition.
  • Page 310 The Basic User Manual provides a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of Voluson®...
  • Page 311 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 10-6 Using a Phantom Refer to the User Manual of the Phantom for information on using a phantom and quality assurance tests.
  • Page 312 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
  • Page 313 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 314 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
  • Page 315 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break.
  • Page 316 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 7.) Follow the test conditions described for respective test points shown in Table 10-14. Table 10-14 Chassis Leakage Current Test Condition...
  • Page 317 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-5-3 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-11.
  • Page 318 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition.
  • Page 319 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-8 Isolated Patient Lead (Sink) Leakage-Isolation Test Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-ISO.
  • Page 320 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the Ultrasound Inspection Certificate.
  • Page 321 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 322 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-9-3 No Meter Probe Adapter Procedure Figure 10-6 Check Without Probe Adapter Follow these steps to test each transducer for leakage current.
  • Page 323 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL 10-7-9-4 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in...
  • Page 324 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Section 10-8 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 325 GE M DRAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL ULTRASOUND INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number: Serial Number: Manufacture Date:...
  • Page 326 GE M RAFT ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe...
  • Page 327 Top Level Architecture, 5-9 in case of System crash or Errors, 7-3 Voluson 730Pro / 730Pro V, 5-10 Delete Voluson 730Pro / 730Pro V Basic, 5-4 Full Backup (Presets, Config. & Appl. Settings), 4-39 Bluetooth Printer Digital Color Printer Connection Scheme, 3-17...
  • Page 328 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Functional Checks, 4-13 ECG Checkout, 4-42 DVD+(R)W Drive Generic Measurements, 4-22 Theory, 5-41 M Mode, 4-11 Mechanical, 4-43 MO-Drive, 4-28 Monitor Display, 4-6...
  • Page 329 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL Installation HP 990cxi, 3-24 CPN6 - Primary Power Supply, 5-42 Installation HP 995c, 3-24 CPN80 - Primary Power Supply, 5-44 List of Abbreviations, 9-2...
  • Page 330 GE M ® 730P / 730P V (BT´04) EDICAL YSTEMS RETZTECHNIK OLUSON KTI105947, R IRECTION EVISION ERVICE ANUAL How to change Video Norm, 3-42 How to enter Date and Time, 3-41 Warnings and Cautions, 1-7 How to enter Hospital Name, 3-41...
  • Page 332 GE Medical Systems: Telex 3797371 P.O. Box 414; Milwaukee, Wisconsin 53201, U.S.A. (Asia, Pacific, Latin America, North America) GE Ultraschall Tel: +49 (0) 212 28 02 208 Deutschland GmbH & Co KG Beethovenstraße 239, Postfach 11 05 60 D-42655 Solingen, Germany...

This manual is also suitable for:

Voluson 730prov

Table of Contents