Reprocessing Procedure - Hitachi EUP-C516 Instruction Manual

Convex array probe
Table of Contents
5.

Reprocessing Procedure

The probe and the magnetic sensor attachment and spacer must be
reprocessed after each use. Refer to the reprocessing instruction
in this chapter.
For reprocessing the biopsy attachment EZU-PA34T and the needle
guide bracket EZU-PA5C1 refer to the instruction manuals of
EZU-PA34T and EZU-PA5C1.
WARNINGS
Limitations on
reprocessing
Transportation
before using
The probe is delivered unsterile. Prior to the
first use, reprocess the probe.
Temperature should not exceed 60°C during
reprocessing.
Probe connector is not water resistant.
The probe is not completely submersible. The
immersible part is shown in Fig.1. The un-immersible
part should be disinfected by wipe disinfection.
The probe should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful not
to damage the sterile pouch or container during
transportation.