Transvaginal probe

Quick Links

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Notes for operators and responsible maintenance personnel 
★  Please read through this Instruction Manual carefully prior to use. 
★  Keep this Instruction Manual together with the system with care to 
make it available anytime. 
 
 
 
 
 
 
 
 
Transvaginal Probe 
EUP‐VV531 
INSTRUCTION MANUAL 
Tokyo, Japan 
© Hitachi, Ltd. 2013, 2017. All rights reserved. 
 
Q1E‐EP1028‐5 
 
 
0123
Table of Contents
loading

Summary of Contents for Hitachi EUP-VV531

  • Page 1           Transvaginal Probe        EUP‐VV531        INSTRUCTION MANUAL              Notes for operators and responsible maintenance personnel  ★  Please read through this Instruction Manual carefully prior to use.  ★  Keep this Instruction Manual together with the system with care to  make it available anytime.                Tokyo, Japan  Q1E‐EP1028‐5    © Hitachi, Ltd. 2013, 2017. All rights reserved.    0123  ...
  • Page 2     Manufacturer:  Hitachi, Ltd.  2‐16‐1, Higashi‐Ueno, Taito‐ku, Tokyo, 110‐0015, Japan  +81‐3‐6284‐3668    http://www.hitachi.com/businesses/healthcare/index.html      European  Representative:  Hitachi Medical Systems GmbH    Otto‐von‐Guericke‐Ring 3 D‐65205 Wiesbaden, Germany         EU Importer:  Hitachi Medical Systems Europe Holding AG  Address:  Sumpfstrasse 13 CH‐6300 Zug, Switzerland      Local Distributor:      (1)  Q1E‐EP1028  ...
  • Page 3   About this manual  This instruction manual shall provide instructions for using, cleaning, and disinfecting the  HITACHI ultrasound probes. It also describes safety considerations, maintenance.      For instructions for operating the main unit, refer to the operation manual for it.    Before  using  the  probe,  thoroughly  read  this  manual  and  keep  this  book  for  future  reference.  If you have any questions concerning the manual, please contact a service support.      The  following  conventions  are  used  throughout  the  manual  to  denote  information  of  special emphasis.   ...
  • Page 4   Graphical Symbols for Use in Labeling of Hitachi Ultrasound Probes  Some  graphical  symbols  that  are  used  in  labeling  of  Hitachi  Ultrasound  Probes  are  compliant with EN980:2008 standard. Refer to the following table about the meanings of  them.    Explanation of Symbol Symbol  Descriptive Content  Manufacturer  Hitachi,Ltd  Company Name and  2‐16‐1, Higashi‐Ueno,taito‐ku,    Address  Tokyo,110‐0015,Japan  Authorized  Hitachi Medical Systems GmbH  Representative in The  Otto‐von‐Guericke‐Ring 3 D‐65205 Wiesbaden,    Germany European Community Store the probe in a cool, dustproof and dark, dry  Keep away from  environment to avoid high temperature and  Sunlight   ...
  • Page 5   Definition of symbol  The following symbol is also used for HITACHI Ultrasound Probes.  Location  Symbol  Definition  Probe  This  instrument  complies  with  Directive  93/42/EEC  relating  to  connector  Medical Device and Directive 2011/65/EU relating to RoHS  Probe  IPX7 mark  IPX7  connector  See section 1.5.  Probe  Type BF APPLIED PART  connector    Probe  General warning sign  connector    Probe  Warning; dangerous voltage  connector    Probe  Caution; Biohazard  connector    Follow  the  instruction  manual  to  operate  this  instrument.  If  not ...
  • Page 6: Table Of Contents

          CONTENTS  Page Introduction ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Features  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Principles of operation ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Intended Use  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Composition  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Construction∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 2 Inspection before Use ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 2 Inspection for Appropriate Connection  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 2 Inspection for Material Surface ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 2 Operation Procedure ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 3 Cleaning and Disinfection ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 4 Point of use (Pre‐cleaning)  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 5 Containment and transportation ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 5 Manual Cleaning and disinfection ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 6 Drying ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 7 Inspection ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 7 Maintenance and Safety Inspection  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 8 Daily check ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 8 Safety Precautions ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 9 Specifications ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 10 Probe ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 10 Suppliers List ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 11 Disposal of the probe  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 11  ...
  • Page 7: Introduction

    3) The  transducer  is  also  used  to  receive  reflected  ultrasonic  waves.  The  transducer  vibrates  mechanically  due  to  the  received  ultrasonic  waves  and  converts  mechanical  vibrations into electric energy. Electric signals are converted to shades of brightness by  brightness modulation to obtain an image.    Intended Use  The Transvaginal Probe EUP‐VV531 is designed for observation and diagnosis mainly of  the following regions by connecting with the HITACHI ultrasound scanner.     Transvaginal     Transrectal      Composition  The probe components of the EUP‐VV531 are as follows: ...
  • Page 8: Construction

      Construction  Immersible part (IPX7) Head  Applied part  Cable  Immersible part:  This part can be immersed in disinfectant  Connector solution and also can be cleaned by water.  Un‐immersible part:    This part should not be immersed in disinfectant  solution and also cannot be cleaned by water.  Fig. 1 External View     Inspection before Use  Prior to use, the probe must be carefully inspected that it is appropriate for use. If not, do  not use the probe and immediately contact a service support.    Inspection for Appropriate Connection  1) Check that the system is correctly operating.    Refer to the instruction manual for the  Main unit.  2) Never  attach  or  connect  unauthorized  devices  or  instruments  on  the  probe,  such  as  unauthorized biopsy attachments. ...
  • Page 9: Operation Procedure

      Operation Procedure  1) Confirm that the probe is disinfected.  2) Connect  the  probe,  operate  the  Main  unit,  and  adjust  the  image  according  to  the  instructions  given in the operation manual for the main unit.  3) Relationship  between  direction  of  the  probe  and  the  right‐left  orientation  mark  on  the  image  is  shown in Fig. 2.  4) Before a diagnosis,  put  a  probe cover to the  tip of  the probe certainly. ...
  • Page 10: Cleaning And Disinfection

      4. Cleaning and Disinfection    The probe and accessory must be reprocessed after each use. Refer to the  reprocessing instruction in this chapter.    ‐ The probe is delivered unsterile. Prior to the first use, reprocess the  probe.  WARNINGS  ‐ Temperature should not exceed 50°C during reprocessing.  ‐ Probe connector is not water resistant.  The probe is not completely submersible. The immersible part is  Limitations on  shown in Fig. 1 The un‐immersible part should be disinfected by wipe  reprocessing  disinfection.      Sterilization is not possible for this probe.  The probe should be packed in a sterile pouch or container to transport  Transportation  from Central Sterile Supply Department (CSSD) to an operating room.  before using  Be careful not to damage the sterile pouch or container during  transportation.  Levels of reprocessing requirements:    Depending on the application of the product and with regard to risk evaluation, the user  has to classify the medical device according to the current Medical Device Directive for  processing of medical devices as uncritical, semi‐critical or critical. Supporting information  concerning this topic is listed in the table below. The user is responsible for correct  classification of the medical device.    Classification  Definition  Processing  Cleaning ...
  • Page 11: Point Of Use (Pre-Cleaning)

      The flowchart of the reprocessing process of this probe is as follows.    Manual cleaning and  disinfection  Point of use  Manual Cleaning Drying  (Pre‐cleaning)  Rinsing after manual  cleaning  Containment and  transportation  Manual Disinfection Rinsing after manual  disinfection      Point of use  Point of use (Pre‐cleaning)    (Pre‐cleaning)  Pre‐cleaning should be done immediately after each use. The procedure is as follows:  1) Remove the protective cover.  2) Clean the probe of all patient’s blood or fluid with running tap water until the surface  of the probe looks visually clean.  3) Wipe  the  whole  surface  of  the  probe  with  gauze  pad  and  remove  superficial  visible  impurities. ...
  • Page 12: Manual Cleaning And Disinfection

      Manual Cleaning and disinfection    Prepare following items before manual cleaning and disinfection:    a) Detergent: Cidezyme (Johnson & Johnson, #2258) or another cleaning agent with  approved material compatibility for this medical device  b) Disinfectant: Cidex OPA (Johnson & Johnson, # 20391) or another disinfectant with  approved material compatibility for this medical device  c) Two tanks, one for cleaning and one for disinfection ‐ optional:  1 additional tank for rinsing with deionized/tap water (sufficient size for immersion of  the immersible part of the probe at full length)  d) Soft, fluff free cloth or single use towel  e) Personal protective equipment (gloves, water repellent protective skirt, face protection  mask or protective glasses, see also instructions of the manufacturer for the detergent  and the disinfectant)    Manual Cleaning:    Prepare the detergent solution in a tank with cold water (please follow the instructions of  the detergent manufacturer regarding application, dilution and contact time).  1) The temperature of the detergent solution should be between 15‐30°C, concentration is  1.6%.  Please  note  the  minimum  contact  time  of  the  detergent  in  the  manufacturer’s  instruction. ...
  • Page 13: Drying

      Manual disinfection:    1) Prepare  the  disinfectant  solution  in  a  tank  with  cold  water  (please  follow  the  instructions  of  the  disinfectant  manufacturer  regarding  application,  concentration,  microbiological efficiency, service life and contact time).  2) Confirm  the  concentration  of  the  disinfectant  before  immersing  the  probe.  Although  ® Cidex  OPA does not need to be diluted, it is recommended to use test strips to verify ...
  • Page 14: Maintenance And Safety Inspection

      Maintenance and Safety Inspection    Daily check  1) Visually check the surface of the probe head, housing and cable for any crack, scratch  or denaturalization. If you find damage, do not use the probe and immediately contact  a service support.    2) After using the probe, it should be cleaned and disinfected according to “4. Cleaning  and  disinfection”,  then  store  it  in  a  cool  and  dark  place  avoid  high  temperature  and  humidity, direct sunlight.    ‐8‐  Q1E‐EP1028  ...
  • Page 15: Safety Precautions

      Safety Precautions  WARNING    Never use the probe if the probe head, housing or cable are cracked or damaged.   Warning  in  case  of  using  probe  covers  which  latex  is  contained  to.    The  latex  may  cause  such  allergic  reactions  as  itching,  rubor,  urticaria,  swelling,  fever,  anhelation,  wheezing, depression of blood pressure, shock and so on.    For the patients suspected  of  latex  allergy,  do  not  use  the  latex‐containing  medical  devices.    If  your  patient  shows ...
  • Page 16: Specifications

      Specifications  Probe  Type:  EUP‐VV531 Transvaginal probe  Center frequency:  6 MHz  Technology:    Convex Array Probe  Dimensions:  See Fig. 4.  Weight:  Approx. 1.0 kg (incl. cable and connector)  Scanning angle:  150  Probe materials:  Bio‐compatible allergy free components  Acoustic output:  According to IEC60601‐2‐37 (See Main Unit manual.)  Applicable system:  Depending on production and upgrade status.    For  detailed information contact a service support.  Classification:  MDD classification IIa.  Cleaning:  Applicable detergents are listed in the suppliers list  Disinfection:  Applicable disinfectants are listed in the suppliers list Operating conditions:    Ambient temperature;  25 ‐ 35C  Contact surface temperature  max. 42C  (temperature of examinee);  Relative humidity;  30 ‐ 85%  Storage conditions: ...
  • Page 17: Suppliers List

      Suppliers List  The products listed below are seriously tested and approved for use with the Transvaginal  Probe EUP‐VV531.    Product name  Manufacturer  Purpose  Klenzyme  Klenzyme  Cleaner  STERANIOS 2%  ANIOS  Disinfectant  Gigasept FF  Schϋlke & Mayr  Disinfectant  CIDEX OPA  Johnson & Johnson Disinfectant  CIDEX  Johnson & Johnson Disinfectant  CIDEX Plus  Johnson & Johnson Disinfectant  Nu‐Cidex  Johnson & Johnson Disinfectant  Stabimed  B. Braun  Disinfectant  ALKAZYME  ALKAPHARM  Cleaning‐ disinfectant Please contact your local distributor for a current version of the ʺDisinfectant/Sterilization  Method Compatibility for Ultrasound Probe and Accessory List”.        Disposal of the probe ...
  • Page 18   28.6  R12.5 28.6 Fig. 4  Dimensions of the probe Unit: mm    ‐12‐  Q1E‐EP1028  ...