Honeywell 854 ATG Series Instruction Manual page 30

Level gauge
Table of Contents
Function
Additional display and function of other switches
switch L
st
Display configuration
1
R display test (blank test / dark test)
U display contrast up
D display contrast down
nd
Display format toggle (formats A - K)
2
U previous display format
D next display format
rd
NOVRAM item view (all items)
3
R fast forward
U previous item
D next item
th
Error list view (item EP)
4
R fast forward
U previous error
D next error
th
System configuration (items: GT, TA, TC, TH, TI, TS, WM)
5
U previous item
D next item
th
Diagnostic view (items: 00, 03, EI, 0S, 0T, EJ, EP, FS, FH, FM, FT, FX, PF, XS, LC)
6
U previous item
D next item
th
Test gauge
7
R test gauge
U lock test
D unlock
The steps of switch L can only be followed in the sequence as given in the table above.
Display scroll
The XPU-2 display can be configured to scroll automatically through all display formats.
This is done by setting item 3Z to enable.
Item
Name
W2=
Protection level 2
3Z=
Toggle display format
EX
Exit
Page 28
Description
Enter protection level 2 password.
With this item, the toggle display format function (display scroll) can be
enabled or disabled. When enabled, all display formats are shown for
approximately 2 seconds.
E :
enable toggle display format
D :
disable toggle display format
Exit protection level.
Operation

Hide quick links:

Table of Contents
loading

Table of Contents