Honeywell Industrial HERMetic Sampler GTX Chem Operation And Service Manual

Honeywell Industrial HERMetic Sampler GTX Chem Operation And Service Manual

Portable closed sampling device
Operation and Service Manual
for Industrial HERMetic Sampler GTX Chem
Portable Closed Sampling Device
Note:
Before using the
instrument please
read this book.
1
50470/SGTX/0909
Industrial HERMetic Sampler GTX Chem
Table of Contents
loading

Summary of Contents for Honeywell Industrial HERMetic Sampler GTX Chem

  • Page 1 Operation and Service Manual for Industrial HERMetic Sampler GTX Chem Portable Closed Sampling Device Note: Before using the instrument please read this book. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 2: Table Of Contents

    PERATING THE RUNNING SAMPLING BOTTLE PERATING THE CARE & MAINTENANCE AFETY WARNING AMPLE CLEANING APE CLEANING APE WIPER ADJUSTMENT OR REPLACEMENT APE REPLACEMENT EARINGS SPECIFICATIONS DRAWINGS & DECLARATION OF CONFORMITY AMPLER ALVES ECLARATION OF ONFORMITY 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 3: Recommendation For Safe Use

    For further guidance refer to International Safety Guide for Oil Tankers and Terminals (ISGOTT), ISBN 10 85609 291 7, Fith Edition 2006, or consult the appropriate Legislative Authority for the installation. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 4: General Information

    The Purchaser shall bear the the warranty is limited to the replacement of cost risk transportation parts or complete goods. defective parts repaired parts supplied in replacement of such defective parts. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 5: Certification

    TS number and description. Refer to section “Drawings”. Some spares might be repairable; in this case send part to any authorized service center or to the factory. In case of urgency replacement units can be available while stocks last. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 6 ..........................................................................................................................................Do you want a quotation before repair is started:..yes / no..... Repaired unit has to be returned to the following address: ................................................................................................................................................................50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 7: Worldwide Service Stations Network

    Ave. Ejército Mexicano 1902 Fax : +52-833-2170190 Col. Loma del Gallo [email protected] 89460 CD. MADERO, TAMPS. MEXICO NETHERLANDS B.V. TECHNISCH BUREAU Tel : +31-10-4114614 UITTENBOGAART Fax : +31-10-4141004 Brugwachter 13 [email protected] NL-3034 KD ROTTERDAM 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 8 LEIGHTON BUZZARD, BEDS LU7 4TX U.S.A / TEXAS HONEYWELL HERMETIC Tel : +1-281-930 1777 4522 Center Street Fax : +1-281-930 1222 DEER PARK, TX 77536 Toll free call in the USA: 1-800-900 1778 [email protected] 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 9: Description

    All these bottle are interchangeable, please refer to § 6.1. For specific application, other bottles exist. For further information, please contact. The sampler is delivered as standard with zone sampling bottle. All other sampling bottles are available as option. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 10: Sampling Principle

    If the instrument is connected to genuine HERMetic valve, grounding is effected through the quick connect coupler and the mating nipple of the valve. No additional grounding strap is necessary. For further information, please refer to §2 “Recommendation for safe use”. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 11: Sampling Method

    The opening of the bottle valve is realized by lowering the sampling bottle until its sitting on the ball of the valve. A pump can be used to accelerate and complete the transfer of the fluid. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 12: Operation

    Nota: Clean the instrument of any excess of liquid after use. Remove the carter winder and clean the storage tube. This cleaning must be done very properly, in particular when corrosive liquids are gauged, such as strong acids or caustic soda for instance. Store the instrument in a dry location. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 13: Operating The Zone Sampling Bottle

    15. Open the HERMetic 2” ball valve no more than 30° to drain any residual liquid back into the tank. 16. After draining close the HERMetic 2” ball valve. 17. Remove the HERMetic Sampler GTX Chem from the HERMetic 2” ball valve. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 14: Operating The Bottom Sampling Bottle

    15. Open the HERMetic 2” ball valve no more than 30° to drain any residual liquid back into the tank. 16. After draining close the HERMetic 2” ball valve. 17. Remove the HERMetic Sampler GTX Chem from the HERMetic 2” ball valve. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 15: Operating The Spot Sampling Bottle

    15. Open the HERMetic 2” ball valve no more than 30° to drain any residual liquid back into the tank. 16. After draining close the HERMetic 2” ball valve. 17. Remove the HERMetic Sampler GTX Chem from the HERMetic 2” ball valve. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 16: Operating The Running Sampling Bottle

    15. Open the HERMetic 2” ball valve no more than 30° to drain any residual liquid back into the tank. 16. After draining close the HERMetic 2” ball valve. 17. Remove the HERMetic Sampler GTX Chem from the HERMetic 2” ball valve. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 17: Care & Maintenance

    Check periodically (at least every 6 months) the continuity of grounding by measuring the electrical resistance between the hook lock (or the sampling bottle) and the quick connect coupler. Resistance should not exceed 100 Ω. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 18: Sample Cleaning

    Check the tape winder for gas tightness (0.3 bar, 4.4 psi) before using again. 7.7 Bearings Bearings are involved in the electrical safety of this device. In case of exchange, use only original spare parts. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 19: Specifications

    II 1 G c IIB T6 (Ta -20 to +80°C) Tape cleaning device Adjustable tape cleaner Available bottles Zone, bottom, spot, running sampling bottles Maintenance modular design / easy exchange of parts Specifications subject to change without notice. 50470/SGTX/0909 Industrial HERMetic Sampler GTX Chem...
  • Page 20: Drawings & Declaration Of Conformity