3.2.1 LCD Display, D-LCC12A

Tools needed
Central Unit, B-CPU6 or B-DISPX and a screwdriver
Visual inspection
Turn off the monitor.
1.
Interface cable
Disconnect and check the LCD display interface cable:
The monitor side connector:
The display side connector:
Check also that the cable itself is intact.
Detach the rear cover by removing the eight screws from the corners.
2.
Check internal parts
Reattach the rear cover.
3.
Check external parts
Functional inspection
Turn on the monitor.
4.
Check the picture
Check that the picture on the LCD display screen is clear and stable.
Check also that the colors are clear.
NOTE: Check that the display screen uses the SVGA mode.
the connector pins are clean and straight and at about the same height
the locking screws are intact
the screw(s) on the casing is tightened properly
the connector pins are clean and intact
the locking screws/claws are intact
all screws are tightened properly
all cables are connected properly
all IC™s that are on sockets are attached properly
there are no loose objects inside the display
the outer cover is intact
the display screen is intact
the block screws for the cable are in place and tightened properly (if installed)
the block screw threads are intact (if installed)
Displays and Controller Boards
Document no. M1137272-02
15