Service Procedures; General Service Information; Service Check; Recommended Tools - GE Datex-Ohmeda S/5 Technical Reference Manual

Anesthesia monitor, critical care monitor
Table of Contents
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
3

Service procedures

3.1 General service information

The field service of the B-CPU6 is limited to replacing the actual CPU board, the
SRAM/Timekeeper battery, or mechanical parts. Faulty CPU board can be returned to GE
Healthcare for repair/exchange.
GE Healthcare is always available for service advice. Please provide the unit serial number, full
type designation and a detailed description of the fault.
WARNING
Only trained personnel with appropriate equipment should perform the
tests and repairs outlined in this section. Unauthorized service may void
warranty of the unit.

3.2 Service check

These instructions include complete procedures for a service check. The service check is
mandatory after any service repair. However, the service check procedures can also be used
for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form
procedures.
The symbol
results of the particular procedure.

3.2.1 Recommended tools

Tool
Command Bar
E-REC
Hemodynamic parameter module
Screwdriver
Visual inspection
Make sure that no cables or modules are connected to the Central Unit. Lift off the display with
its stand, if placed on the Central Unit.
CPU board
1.
Device plates
Check that the CPU board rear panel is clean and intact. Check that both CPU and
software device plates are intact.
10
.
Document no
M1125637-04
("APPENDIX
in the instructions indicates that the check form contains space to record the
Order No.
A:") which may be used when performing the
Notes