Chlorine Readings Too Low; Total Chlorine Measurement And Calibration Problems; Ozone Measurement And Calibration Problems; Ozone Zero Current Outside Range - Emerson Rosemount5081 Instruction Manual

Explosion-proof transmitter
Table of Contents
Diagnostics and troubleshooting

Chlorine readings too low

1.
2.
3.
8.4.6

Total chlorine measurement and calibration problems

Refer to the instruction manual for the SCS921 for a complete troubleshooting guide.
8.4.7

Ozone measurement and calibration problems

Problem
Zero current is substantially outside the
range -10 to 10 nA.
Zero reading is unstable.
Sensor current during calibration is sub-
stantially less than about 350 nA/ppm at
25 °C (77 °F).
Process readings are erratic.
Readings drift.
Sensor does not respond to changes in
ozone level.
Ozone readings are too low.

Ozone zero current outside range

The ozone zero current is substantially outside the range -10 to 10 nA.
1.
2.
106
Was the sample tested as soon as it was taken? Chlorine solutions are unstable. Test
the sample immediately after collecting it. Avoid exposing the sample to sunlight.
Low readings can be caused by zeroing the sensor before the residual current has
reached a stable minimum value. Residual current is the current the sensor
generates even when no chlorine is in the sample. Because the residual current is
subtracted from subsequent measured currents, zeroing before the current is a
minimum can lead to low results. See
Sensor response depends on flow. If the flow is too low, readings will be low and flow
sensitive. Verify that the flow past the sensor equals or exceeds the minimum value.
See the sensor instruction manual for recommended flows.Verify that the chlorine
sensor is installed in the correct flow cell. See Figure and Figure. Verify that the liquid
level in the constant head sampler is level with the central overflow tube and that
excess sample is flowing down the tube. If necessary, disassemble and clean the
overflow sampler. See Section.
Is the sensor properly wired to the transmitter?
Is the zero solution ozone free? Test the zero solution for ozone level. The
concentration should be less than 0.02 ppm.
Oxygen readings too low
See Section
Ozone zero current outside range
Ozone zero reading unstable
Sensor current less than 350 nA/ppm
Ozone process readings erratic
Ozone readings drift
Sensor not responding to ozone changes
Ozone readings too low
for more information.