Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

15 MIS (Master Interconnect System)

Innova system interconnect cables are described in MIS (Master Interconnect System) documents.
These documents specify all interconnections between components within the system.
Reference: For specific Vascular system interconnect maps and connection details, refer to the
following
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ MIS Map
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ MIS Charts
General Guidelines
Innova System introduce a new system interconnect with a star distribution for all cables from the
technical area. The cables are shipped on spools to create cable groups. Cable group 1 for Exam
room and cable group 2 for control room. The cable group shall be put in place during the same
action. The cables are routed in the same duct.
The HV cables could be pulled separately.
Chapter 6 Electrical Connections
Direction 5160944-13-1EN, Revision 2
199