Testing - Honeywell Midas S2 User Manual

Fixed single-point extractive gas transmitter
Hide thumbs Also See for Midas S2:
Table of Contents

3.5 Testing

From the Test Mode, you can test Bump, Inhibit, 4-20mA, Alarm/Fault simulation, Relay, Line
Integrity Test, and Blocking test.
1. Go to the Monitoring mode.
2. Press
3. Press
4. Optional Step: Press
Table 3-4: Test mode menu
Option
Bump
Inhibit
4-20mA
Alarm/Fault
simulation
Relay
LIT (optional)
Blocking test
NOTE: All test items display results only within the test mode, and the device goes back to the
normal status when exiting the test mode except LIT test. If the LIT test fails, the monitoring
mode shows the LIT fail and fault indication.
Midas S2
and select Test Mode, and then press
or
to scroll the options, and then press
to cancel and return one step back.
Operations you can Test from the selected option
l
Gas bump test . For bump test details, please refer to the Topic
6.1:Calibration & Bump test
Inhibit ON/OFF
l
Inhibit time
l
Inhibit current level
l
l
Force 1mA - 21mA output of each channel
Alarm 1
l
Alarm 2
l
Maintenance Fault
l
Instrument Fault
l
l
Relays 1, 2 and Fault
l
Relay energized/de-energized in normal condition
LIT Test
l
to enter the menu.
to accept it.