Analog Bus Timing - HP 8560E Service Manual

Spectrum analyzers
Table of Contents
ADC/Interface Section
A3 Assembly Control Circuits
12.If any address or data line is low, press
analyzer off and disconnect the W2 control cable from A3J2. Press
to turn spectrum analyzer on. Ignore any error messages.
LINE
13.Check that address lines A0 through A7 and data lines D0 through
D7 are all high. If all address and data lines are high, suspect a fault
either in W2 or one of the other four assemblies which connect to
W2.
14.If any address or data line is low, check the appropriate input of
either U405 (data lines) or U406 (address lines).
15.If a data line input is stuck low, check the data bus buffer. If an
address line input is stuck low, check A3W1 and the A2 controller
assembly.
16.If the appropriate input is high or toggling between high and low,
suspect a failure in either U405 (data lines) or U406 (address lines).
17.Remove jumpers.

Analog Bus Timing

Refer to function block P of A3 Interface Assembly Schematic Diagram
(sheet 4 of 6) in the
HP 8560 E-Series Spectrum Analyzer Component Level
.
Information
Analog bus timing (ABT) generates the strobes for the A4, A5, A14, and
A15 assemblies. The A14 frequency control assembly also requires a
qualifier for its strobe, LVFC_ENABLE. A3U400 and A3U414 provide a
2 µs delay between the time HANA_BUS goes high and the enable line
to demultiplexer A3U407 goes low.
1. Press
on the HP 8560E/EC and set the controls as follows:
PRESET
Center frequency ..................................................... 300 MHz
Span ......................................................................... 100 MHz
2. Check that A3U407 pin 1 goes low approximately 2 µs after
HANA_BUS (A3U400 pin 3) goes high.
3. If HANA_BUS is absent, check for pulses on ABT A3U505 pin 2 and
IA10 (A3U505 pin 5).
4. If A3U407 pin 1 is not delayed 2 µs from HANA_BUS, check for the
presence of the 1 MHz HBADC_CLK0.
5. If A3U407 pin 1 is not delayed 2 µs from HANA_BUS and
HBADC_CLK0 is correct, suspect a fault in either A3U414 or
A3U400.
6. Press
and set the controls as follows:
PRESET
Span ................................................................................. 0 Hz
Trigger ...................................................................... SINGLE
386
to turn spectrum
LINE
Chapter 7
Table of Contents
loading

This manual is also suitable for:

8560ec

Table of Contents