Battery Problem (718); Model Number Error (719); System Errors (750 To 759) - HP 8560E Service Manual

Spectrum analyzers
Table of Contents
General Troubleshooting
System Analyzer Programming Errors (100 to 150)
713 RAM U301 Checksum error of system RAM A2U301.
714 RAM U300 Checksum error of system RAM A2U300.
715 RAM U305 Checksum error of system RAM A2U305.
716 RAM U304 Checksum error of system RAM A2U304.
Microprocessor Error (717)
Microprocessor not fully operational. Refer to Chapter
717 BAD uP
9, "Controller Section."

Battery Problem (718)

If STATE or TRACE data is found to be corrupt, the
processor tests the display RAMs and the program
RAMs containing the STATE information. If the RAMs
are working properly, this error message is generated.
To check the BT1 battery and the battery backup
circuitry, refer to "STATE and TRACE Storage
Problems" in Chapter 9.
718 BATTERY? Nonvolatile RAM not working; check battery BT1.
This error can also be generated if the battery has been
disconnected then reconnected. If this is the cause,
cycling power clears the error.

Model Number Error (719)

If this error occurs, return the instrument to a service
center for repair.
719 MODEL #? Could not read ID string from EEROM A2U501.

System Errors (750 to 759)

These errors often require troubleshooting the A2
controller and A3 interface assemblies.
Hardware/firmware interaction, zero divide. Check for
750 SYSTEM
other errors.
Hardware/firmware interaction, floating point overflow.
751 SYSTEM
Check for other errors.
Hardware/firmware interaction, floating point
752 SYSTEM
underflow. Check for other errors.
Hardware/firmware interaction, log error. Check for
753 SYSTEM
other errors.
Hardware/firmware interaction, integer overflow.
754 SYSTEM
Check for other errors.