Figure 2-27 16 Mhz Pll Adjustment Location - HP 8560E Service Manual

Spectrum analyzers
Table of Contents
Adjustment Procedures
19. 16 MHz PLL Adjustment
Figure 2-27
16 MHz PLL Adjustment Location
4. Press
spectrum analyzer power-on adjustments have completed.
5. Set the microwave frequency counter as follows:
6. On the HP 8560E/EC spectrum analyzer, press
PANEL
128
to turn the spectrum analyzer on. Wait until the
LINE
Sample rate ........................................................................ Fully counterclockwise
10 Hz — 500 MHz/500 MHz — 26.5 GHz Switch.................10 Hz — 500 MHz
50 Ω/1 MΩ Switch ............................................................ 1 MΩ
, and
.
10 MHz EXT
,
AUX CTRL
REAR
Chapter 2
Table of Contents
loading

This manual is also suitable for:

8560ec

Table of Contents